Chiral separation of synthetic vicinal diol compounds by capillary zone electrophoresis with borate buffer and beta-cyclodextrin as buffer additive.

نویسندگان

  • Yan Zhao
  • Xingbin Yang
  • Ru Jiang
  • Xiaoli Sun
  • Wenmin Liu
  • Shengyong Zhang
چکیده

The investigation on capillary electrophoretic enantioseparation of six synthetic compounds containing vicinal diol groups has been undertaken to acquire the optimum conditions using native beta-cyclodextrin (beta-CD) as chiral selector and borate as a background electrolyte. The separation was carried out in an uncoated capillary (58.5 cm x 75 microm i.d., effective length 48.5 cm) and the effects of several important factors were investigated in detail. The results showed that beta-CD as a chiral selector exhibited good enantioselectivity and that the enantioseparation was greatly influenced by the structure of the diols, the borate concentration and the buffer pH. The optimum performance was obtained for the chiral vicinal diols under the conditions of 200 mM borate buffer of pH 9.8 containing 1.7% beta-CD at an applied voltage of 15 kV and a capillary temperature of 20 degrees C. Under the conditions, four diols were baseline separated with fast analysis time and the good theoretical plate numbers (above 10 x 10(4)) and favorable migration-time reproducibilities (RSDs below 3.0%) were obtained. The separation results were satisfactory.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Application of Guanidine Hcl to Improve Enantioseparation of a Model Basic Drug, Cetirizine, By Capillary Electrophoresis Using Sulfated Β-Cyclodextrin

A common approach in resolving enantiomers of chiral basic drugs by capillary electrophoresis (CE) is to use cyclodextrins (especially their anionic derivatives) as chiral selector in the acidic buffer (pH ≤ 3) in normal or reversed (carrier) mode. Then, some organic modifiers are added to the buffer solution if the resolution is not satisfactory. In case of cetirizine (CTN), applying the same ...

متن کامل

Application of Guanidine Hcl to Improve Enantioseparation of a Model Basic Drug, Cetirizine, By Capillary Electrophoresis Using Sulfated Β-Cyclodextrin

A common approach in resolving enantiomers of chiral basic drugs by capillary electrophoresis (CE) is to use cyclodextrins (especially their anionic derivatives) as chiral selector in the acidic buffer (pH ≤ 3) in normal or reversed (carrier) mode. Then, some organic modifiers are added to the buffer solution if the resolution is not satisfactory. In case of cetirizine (CTN), applying the same ...

متن کامل

Improvement of Capillary Electrophoretic Enantioseparation of Fluoxetine by a Cationic Additive

One of the problems encountered in CE separations of basic compounds is the adsorption of analytes onto the negatively charged capillary wall which could lead to poor repeatability of migration time and peak area. Additionally, separation of enantiomers of chiral of basic drugs is commonly carried out in low pH buffer which contributes to strong ionic interaction of the cationic drug ions with ...

متن کامل

Improvement of Capillary Electrophoretic Enantioseparation of Fluoxetine by a Cationic Additive

One of the problems encountered in CE separations of basic compounds is the adsorption of analytes onto the negatively charged capillary wall which could lead to poor repeatability of migration time and peak area. Additionally, separation of enantiomers of chiral of basic drugs is commonly carried out in low pH buffer which contributes to strong ionic interaction of the cationic drug ions with ...

متن کامل

Improvement of Electrophoretic Enantioseparation of Amlodipine by Polybrene

   In chiral and non-chiral electrophoretic resolution of basic drugs, adsorption of analytes to negatively charged capillary wall could lead to poor repeatability of migration time and peak area. In addition, chiral resolutions of basic drugs are commonly performed in low pH buffers. Therefore, longer analysis time due to suppression of electroosmotic flow (EOF) is another dilemma. In this wor...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Analytical sciences : the international journal of the Japan Society for Analytical Chemistry

دوره 22 5  شماره 

صفحات  -

تاریخ انتشار 2006