The Sensitivity of Clinical Outcomes to Centration on the Light-Constricted Pupil for a Shape-Changing Corneal Inlay.

نویسندگان

  • Enrique Barragán-Garza
  • Douglas D Koch
  • Luis G Vargas
  • Alan Lang
  • Adam Roy
چکیده

PURPOSE To assess the clinically acceptable range of inlay decentration with respect to the light-constricted pupil center and the coaxially sighted corneal light reflex (CSCLR) for an inlay (Raindrop Near Vision Inlay; ReVision Optics, Inc., Lake Forest, CA) that reshapes the anterior corneal surface. METHODS In this retrospective, observational cohort study of 115 patients with emmetropic or low hyperopic presbyopia who were implanted with a shape-changing corneal inlay, visual acuity, task performance (in good and dim light), reports of halos and glare, and satisfaction data were collected from the preoperative and 3-month postoperative examinations. Inlay centration with respect to the pupil center and CSCLR was determined from the center of the inlay effect derived from iTrace (Tracey Technologies, Houston, TX) wavefront measurements. Multivariate regression models assessed the influence of inlay position on visual outcomes. RESULTS On average, monocular uncorrected near visual acuity (UNVA) improved 4.9 ± 1.7 lines in the treated eye, with no loss in binocular distance vision. Eighty-three percent of implants were centered radially within 0.5 mm of the pupil center. Multivariate analysis of decentration with respect to both the pupil center and CSCLR revealed no significant interaction with the above clinical outcomes, with the exception of UNVA in the treated eye (all P > .05, α = 0.05). For decentration of less than 0.75 mm, the change in UNVA was less than 1 line. CONCLUSIONS Distance and near visual acuity, task performance, severity of halos and glare, and satisfaction were independent of radial decentration of the Raindrop Near Vision Inlay of less than 0.75 mm from the light-constricted pupil. [J Refract Surg. 2018;34(3):164-170.].

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Vignetting and Field of View with the KAMRA Corneal Inlay

PURPOSE To evaluate the effect of the KAMRA corneal inlay on the retinal image brightness in the peripheral visual field. METHODS A KAMRA inlay was "implanted" into a theoretical eye model in a corneal depth of 200 microns. Corneal radius was varied to a steep, normal, and flat (7.37, 7.77, and 8.17 mm) version keeping the proportion of anterior to posterior radius constant. Pupil size was va...

متن کامل

Centration axis in refractive surgery

The human eye is an asymmetric optical system and the real cornea is not a rotationally symmetrical volume. Each optical element in the eye has its own optical and neural axes. Defining the optimum center for laser ablation is difficult with many available approaches. We explain the various centration approaches (based on these reference axes) in refractive surgery and review their clinical out...

متن کامل

Comparison of myopic LASIK centered on the coaxially sighted corneal light reflex or line of sight.

PURPOSE To compare refractive outcomes of myopic LASIK with centration on the coaxially sighted corneal light reflex (CSCLR) to centration on the center of the pupil (line of sight [LOS]). METHODS The NIDEK CXIII excimer laser was used to treat 268 eyes with centration on the CSCLR (CSCLR group) and 288 eyes with centration on the LOS (LOS group). For the CSCLR group, the laser ablation was d...

متن کامل

Location of achromatizing pupil position and first Purkinje reflection in a normal population.

PURPOSE Quality of vision in patients who have undergone corneal refractive surgery depends upon the optimal centration of the procedures used. The center of the pupil is used as a reference point in some corneal ablation procedures. The achromatic axis would be a more sensible option from an optical point of view, but it is not as readily detectable. As an alternative, other refractive techniq...

متن کامل

Photic sensitivity ranges of hamster pupillary and circadian phase responses do not overlap.

Mammalian retinal photoreceptors form an irradiance detection system that drives many nonvisual responses to light such as pupil reflex and resetting of the circadian clock. To understand the role of pupil size in circadian light responses, pupil diameter was pharmacologically manipulated and the effect on behavioral phase shifts at different irradiance levels was studied in the Syrian hamster....

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of refractive surgery

دوره 34 3  شماره 

صفحات  -

تاریخ انتشار 2018