Use of esophageal balloon pressure-volume curve analysis to determine esophageal wall elastance and calibrate raw esophageal pressure: a bench experiment and clinical study
نویسندگان
چکیده
BACKGROUND Accurate measurement of esophageal pressure (Pes) depends on proper filling of the balloon. Esophageal wall elastance (Ees) may also influence the measurement. We examined the estimation of balloon-surrounding elastance in a bench model and investigated a simplified calibrating procedure of Pes in a balloon with relatively small volume. METHODS The Cooper balloon catheter (geometric volume of 2.8 ml) was used in the present study. The balloon was progressively inflated in different gas-tight glass chambers with different inner volumes. Chamber elastance was measured by the fitting of chamber pressure and balloon volume. Balloon pressure-volume (P-V) curves were obtained, and the slope of the intermediate linear section was defined as the estimated chamber elastance. Balloon volume tests were also performed in 40 patients under controlled ventilation. The slope of the intermediate linear section on the end-expiratory esophageal P-V curve was calculated as the Ees. The balloon volume with the largest Pes tidal swing was defined as the best volume. Pressure generated by the esophageal wall during balloon inflation (Pew) was estimated as the product of Ees and best volume. Because the clinical intermediate linear section enclosed filling volume of 0.6 to 1.4 ml in each of the patient, we simplified the estimation of Ees by only using parameters at these two filling volumes. RESULTS In the bench experiment, bias (lower and upper limits of agreement) was 0.5 (0.2 to 0.8) cmH2O/ml between the estimated and measured chamber elastance. The intermediate linear section on the clinical and bench P-V curves resembled each other. Median (interquartile range) Ees was 3.3 (2.5-4.1) cmH2O/ml. Clinical best volume was 1.0 (0.8-1.2) ml and ranged from 0.6 to 1.4 ml. Estimated Pew at the best volume was 2.8 (2.5-3.5) cmH2O with a maximum value of 5.2 cmH2O. Compared with the conventional method, bias (lower and upper limits of agreement) of Ees estimated by the simple method was - 0.1 (- 0.7 to 0.6) cmH2O/ml. CONCLUSIONS The slope of the intermediate linear section on the balloon P-V curve correlated with the balloon-surrounding elastance. The estimation of Ees and calibration of Pes were feasible for a small-volume-balloon. TRIAL REGISTRATION Identifier NCT02976844 . Retrospectively registered on 29 November 2016.
منابع مشابه
In vivo calibration of esophageal pressure in the mechanically ventilated patient makes measurements reliable
BACKGROUND Esophageal pressure (Pes) can provide information to guide mechanical ventilation in acute respiratory failure. However, both relative changes and absolute values of Pes can be affected by inappropriate filling of the esophageal balloon and by the elastance of the esophagus wall. We evaluated the feasibility and effectiveness of a calibration procedure consisting in optimization of b...
متن کاملOptimal esophageal balloon volume for accurate estimation of pleural pressure at end-expiration and end-inspiration: an in vitro bench experiment
BACKGROUND Esophageal pressure, used as a surrogate for pleural pressure, is commonly measured by air-filled balloon, and the accuracy of measurement depends on the proper balloon volume. It has been found that larger filling volume is required at higher surrounding pressure. In the present study, we determined the balloon pressure-volume relationship in a bench model simulating the pleural cav...
متن کاملSuccessful Use of Biliary Duct Balloon Dilator in Repairing Post-Surgical Esophageal Stricture in premature infant
Congenital esophageal atresia needs to be surgically corrected as soon as possible. Some will present with post surgical stricture that needs to be dilated by a balloon dilator. This procedure is difficult in premature born due to infant esophageal size and availability of equipments. We herby report the first case in a premature infant with post surgical esophageal stricture of esophageal atre...
متن کاملRelationship between esophageal clinical symptoms and manometry findings in patients with esophageal motility disorders: a cross-sectional study
Background: Manometry is the gold-standard diagnostic test for motility disorders in the esophagus. The development of high-resolution manometry catheters and software displays of manometry recordings in color-coded pressure plots have changed the diagnostic assessment of esophageal disease. The diagnostic value of particular esophageal clinical symptoms among patients suspected of esophageal m...
متن کاملمانومتری با قدرت تفکیک بالا در افتراق اختلالات حرکتی مری
Background: High-resolution manometer (HRM) of the esophagus has become the main diagnostic test in the evaluation of esophageal motility disorders. The development of high-resolution manometry catheters and software displays of manometry recordings in color-coded pressure plots have changed the diagnostic assessment of esophageal disease. The first step of the Chicago classification described ...
متن کامل