Intraoperative Defibrillation Testing of Subcutaneous Implantable Cardioverter‐Defibrillator Systems—A Simple Issue?
نویسندگان
چکیده
BACKGROUND The results of the recently published randomized SIMPLE trial question the role of routine intraoperative defibrillation testing. However, testing is still recommended during implantation of the entirely subcutaneous implantable cardioverter-defibrillator (S-ICD) system. To address the question of whether defibrillation testing in S-ICD systems is still necessary, we analyzed the data of a large, standard-of-care prospective single-center S-ICD registry. METHODS AND RESULTS In the present study, 102 consecutive patients received an S-ICD for primary (n=50) or secondary prevention (n=52). Defibrillation testing was performed in all except 4 patients. In 74 (75%; 95% CI 0.66-0.83) of 98 patients, ventricular fibrillation was effectively terminated by the first programmed internal shock. In 24 (25%; 95% CI 0.22-0.44) of 98 patients, the first internal shock was ineffective and further internal or external shock deliveries were required. In these patients, programming to reversed shock polarity (n=14) or repositioning of the sensing lead (n=1) or the pulse generator (n=5) led to successful defibrillation. In 4 patients, a safety margin of <10 J was not attained. Nevertheless, in these 4 patients, ventricular arrhythmias were effectively terminated with an internal 80-J shock. CONCLUSIONS Although it has been shown that defibrillation testing is not necessary in transvenous ICD systems, it seems particular important for S-ICD systems, because in nearly 25% of the cases the primary intraoperative test was not successful. In most cases, a successful defibrillation could be achieved by changing shock polarity or by optimizing the shock vector caused by the pulse generator or lead repositioning.
منابع مشابه
Defibrillation Testing of the Implantable Cardioverter Defibrillator: When, How, and by Whom?
The implantable cardioverter-defibrillator (ICD) has become an integral part of treatment for a variety of patients with symptomatic, or at risk for, ventricular tachyarrhythmias. The ICD's effectiveness is attributed to its ability to promptly detect and terminate ventricular tachycardia (VT) and fibrillation (VF). The clinical trials that established the positive role of ICD therapy were base...
متن کاملSubcutaneous Implantable Cardioverter Defibrillators Implantation Without Defibrillation Threshold Testing: A Single Center Experience
Background Subcutaneous implantable cardioverter defibrillator (S-ICD) system has been proven to be an effective therapy for prevention of sudden cardiac death (SCD) in selected patients. Although the Shockless IMPLant Evaluation (SIMPLE) trial has shown that defibrillation threshold (DFT) testing is not necessary for transvenous ICD (TV-ICD) systems, it is still recommended for S-ICD systems. ...
متن کاملSuccessful intermuscular implantation of subcutaneous implantable cardioverter defibrillator in a Japanese patient with pectus excavatum
The entirely subcutaneous implantable cardioverter-defibrillator (ICD) system was developed to provide a life-saving defibrillation therapy that does not affect the heart and vasculature. The subcutaneous ICD is preferred over the transvenous ICD for patients with a history of recurrent infection presenting major life-threatening rhythms. In this case report, we describe the first successful in...
متن کاملIs there a need for routine testing of ICD defibrillation capacity? Results from more than 1000 studies.
AIMS Benefits and complications of postoperative implantable cardioverter-defibrillator tests are controversial matters. This study sought to assess the necessity of defibrillation function tests after implantation. METHODS AND RESULTS We retrospectively analysed 1007 implantable cardioverter-defibrillator tests in 587 systems and 556 patients. Nine hundred and thirty implantable cardioverter...
متن کاملSubcutaneous implantable cardioverter-defibrillator placement in a patient with a preexisting transvenous implantable cardioverter-defibrillator
Introduction The subcutaneous implantable cardioverter-defibrillator (S-ICD) is a safe and effective treatment for prevention of sudden cardiac death. It is typically not used in patients who require additional therapies conferred by transvenous implantable cardioverter-defibrillator (TV-ICD) systems, such as bradycardia pacing or antitachycardia pacing (ATP) for ventricular tachycardia (VT). D...
متن کامل