Inductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations).
نویسندگان
چکیده
BACKGROUND Implantable cardioverter-defibrillators and cardiac resynchronization therapy defibrillators have relied on multiple ventricular fibrillation (VF) induction/defibrillation tests at implantation to ensure that the device can reliably sense, detect, and convert VF. The ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations) is the first large, multicenter, prospective trial comparing vulnerability safety margin testing versus defibrillation safety margin testing with a single VF induction/defibrillation. METHODS AND RESULTS A total of 426 patients receiving an implantable cardioverter-defibrillator or cardiac resynchronization therapy defibrillator underwent vulnerability safety margin or defibrillation safety margin screening at 14 J in a randomized order. After this, patients underwent confirmatory testing, which required 2 VF conversions without failure at < or = 21 J. Patients who passed their first 14-J and confirmatory tests, irrespective of the results of their second 14-J test, had their devices programmed to a 21-J shock for ventricular tachycardia (VT) or VF > or = 200 bpm and were followed up for 1 year. Of 420 patients who underwent 14-J vulnerability safety margin screening, 322 (76.7%) passed. Of these, 317 (98.4%) also passed 21-J confirmatory tests. Of 416 patients who underwent 14-J defibrillation safety margin screening, 343 (82.5%) passed, and 338 (98.5%) also passed 21-J confirmatory tests. Most clinical VT/VF episodes (32 of 37, or 86%) were terminated by the first shock, with no difference in first shock success. In all observed cases in which the first shock was unsuccessful, subsequent shocks terminated VT/VF without complication. CONCLUSIONS Although spontaneous episodes of fast VT/VF were limited, there was no difference in the odds of first shock efficacy between groups. Screening with vulnerability safety margin or defibrillation safety margin may allow for inductionless or limited shock testing in most patients.
منابع مشابه
Automated vulnerability testing identifies patients with inadequate defibrillation safety margin.
BACKGROUND Implantable cardioverter-defibrillator system efficacy is tested at implant by induction of ventricular fibrillation (VF). Defibrillation safety margin can be assessed without VF induction using upper limit of vulnerability methods, but these methods have required manual determination of T-wave timing. METHODS AND RESULTS To test the feasibility of an inductionless system of implan...
متن کاملImplantation of cardioverter defibrillators without induction of ventricular fibrillation.
BACKGROUND The upper limit of vulnerability (ULV) is the weakest shock at which ventricular fibrillation (VF) is not induced by a T-wave shock. This study tested the hypothesis that a vulnerability safety margin based on the ULV can be used as an implantable cardioverter-defibrillator implantation criterion. METHODS AND RESULTS Implantable cardioverter-defibrillators were implanted in 80 pati...
متن کاملICD Shock, Not Ventricular Fibrillation, Causes Elevation of High Sensitive Troponin T after Defibrillation Threshold Testing—The Prospective, Randomized, Multicentre TropShock-Trial
BACKGROUND The placement of an implantable cardioverter defibrillator (ICD) has become routine practice to protect high risk patients from sudden cardiac death. However, implantation-related myocardial micro-damage and its relation to different implantation strategies are poorly characterized. METHODS A total of 194 ICD recipients (64±12 years, 83% male, 95% primary prevention of sudden cardi...
متن کاملSystematic Review of Defibrillation Threshold Testing at De Novo Implantation.
BACKGROUND Recent results from the largest multicenter randomized trial (Shockless IMPLant Evaluation [SIMPLE]) on defibrillation threshold (DFT) testing suggest that while shock testing seems safe, it does not reduce the risk of failed shocks or prolong survival. A contemporary systematic review of DFT versus no-DFT testing at the time of implantable cardioverter-defibrillator implantation was...
متن کاملSuccess of Device-Based Methods for Atrial Tachyarrhythmia Induction during Implantation of an Implantable Cardioverter Defibrillator with Atrial Therapy
In addition to standard ventricular therapy, cardioversion using an implantable cardioverter defibrillator (ICD) with atrial therapy is one of the many new options available for treating atrial tachyarrhythmias. New data suggest that shock efficacy for AF depends on identifying atrial defibrillation thresholds (DFT) and programming the first atrial shock to at least two times the atrial DFT [1]...