Quantitative evaluation of the performance of different deformable image registration algorithms in helical, axial, and cone‐beam CT images using a mobile phantom

نویسندگان

  • Imad Ali
  • Nesreen Alsbou
  • Justin Jaskowiak
  • Salahuddin Ahmad
چکیده

The goal of this project is to investigate quantitatively the performance of different deformable image registration algorithms (DIR) with helical (HCT), axial (ACT), and cone-beam CT (CBCT). The variations in the CT-number values and lengths of well-known targets moving with controlled motion were evaluated. Four DIR algorithms: Demons, Fast-Demons, Horn-Schunck and Lucas-Kanade were used to register intramodality CT images of a mobile phantom scanned with different imaging techniques. The phantom had three water-equivalent targets inserted in a low-density foam with different lengths (10-40 mm) and moved with adjustable motion amplitudes (0-20 mm) and frequencies (0-0.5 Hz). The variations in the CT-number level, volumes and shapes of these targets were measured from the spread-out of the CT-number distributions. In CBCT, most of the DIR algorithms were able to produce the actual lengths of the mobile targets; however, the CT-number values obtained from the DIR algorithms deviated from the actual CT-number of the targets. In HCT, the DIR algorithms were successful in deforming the images of the mobile targets to the images of the stationary targets producing the CT-number values and lengths of the targets for motion amplitudes <20 mm. Similarly in ACT, all DIR algorithms produced the actual CT-number values and lengths of the stationary targets for low-motion amplitudes <15 mm. The optical flow-based DIR algorithms such as the Horn-Schunck and Lucas-Kanade performed better than the Demons and Fast-Demons that are based on attraction forces particularly at large motion amplitudes. In conclusion, most of the DIR algorithms did not reproduce well the CT-number values and lengths of the targets in images that have artifacts induced by large motion amplitudes. The deviations in the CT-number values and variations in the volume of the mobile targets in the deformed CT images produced by the different DIR algorithms need to be considered carefully in the treatment planning for accurate dose calculation dose coverage of the tumor, and sparing of critical structures.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Study on Robustness of Various Deformable Image Registration Algorithms on Image Reconstruction Using 4DCT Thoracic Images

Background: Medical image interpolation is recently introduced as a helpful tool to obtain further information via initial available images taken by tomography systems. To do this, deformable image registration algorithms are mainly utilized to perform image interpolation using tomography images.Materials and Methods: In this work, 4DCT thoracic images of five real patients provided by DI...

متن کامل

Quantitative SPECT and planar 32P bremsstrahlung imaging for dosimetry purpose –An experimental phantom study

Background: In this study, Quantitative 32P bremsstrahlung planar and SPECT imaging and consequent dose assessment were carried out as a comprehensive phantom study to define an appropriate method for accurate Dosimetry in clinical practice. Materials and Methods: CT, planar and SPECT bremsstrahlung images of Jaszczak phantom containing a known activity of 32P were acquired. In addition, Phanto...

متن کامل

A New Approach for Quantitative Evaluation of Reconstruction Algorithms in SPECT

ABTRACT Background: In nuclear medicine, phantoms are mainly used to evaluate the overall performance of the imaging systems and practically there is no phantom exclusively designed for the evaluation of the software performance.  In this study the Hoffman brain phantom was used for quantitative evaluation of reconstruction techniques. The phantom is modified to acquire t...

متن کامل

Evaluation of deformable image registration in HDR gynecological brachytherapy

Introduction: In brachytherapy, as in external radiotherapy, image-guidance plays an important role. For GYN treatments it is standard to acquire at least CT images and preferably MR images prior to each treatment and to calculate the dose of the day on each set of images. Then, the dose to the target and to the organs at risk (OAR) is calculated with worst case scenario from I...

متن کامل

Evaluation of the gray level in CBCT systems and its relationship with HU in CT Scanners

Introduction: Cone-beam CT (CBCT) is an imaging system which offers three-dimensional (3D), multiplanar images and has many advantages over computed tomography (CT) including shorter acquisition times for the resolution desired in dentistry, lower radiation dose to the patient, reasonable price and higher spatial resolution but CBCT scanners are unable to display actual Hounsf...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 19  شماره 

صفحات  -

تاریخ انتشار 2018