Image guidance during breast radiotherapy: a phantom dosimetry and radiation-induced second cancer risk study
نویسنده
چکیده
Imaging procedures utilised for patient position verification during breast radiotherapy can add a considerable dose to organs surrounding the target volume on top of therapeutic scatter dose. This study investigated the dose from a breast kilovoltage cone-beam CT (kV-CBCT), a breast megavoltage fan-beam CT (MV-FBCT), and a TomoDirect breast treatment. Thermoluminescent dosimeters placed within a female anthropomorphic phantom were utilised to measure the dose to various organs and tissues. The contralateral breast, lungs and heart received 0.40 cGy, 0.45 cGy and 0.40 cGy from the kV-CBCT and 1.74 cGy, 1.39 cGy and 1.73 cGy from the MV-FBCT. In comparison to treatment alone, daily imaging would increase the contralateral breast, contralateral lung and heart dose by a relative 12%, 24% and 13% for the kV-CBCT, and 52%, 101% and 58% for the MV-FBCT. The impact of the imaging dose relative to the treatment dose was assessed with linear and linear-quadratic radiation-induced secondary cancer risk models for the contralateral breast. The additional imaging dose and risk estimates presented in this study should be taken into account when considering an image modality and frequency for patient position verification protocols in breast radiotherapy.
منابع مشابه
Out-Of-Field Dose Measurement and Second Cancer-Risk Estimation Following External Beam Radiotherapy and Brachytherapy for Cervical Cancer Treatment: A Phantom Study
Introduction: The present study aimed to measure the scatter and leakage dose received by out-of-field organs while delivering Radiotherapy (RT) treatment of cervical cancer. Moreover, this study estimated the risk of second cancer (SC). The doses to out-of-field organs were measured using a lithium fluoride (TLD 100) dosimeter while delivering External Beam Radiothera...
متن کاملA Dosimetric Evaluation of Organs at Risk in Prostate Radiation Therapy using a MAGIC Gel Dosimeter
Introduction: Multiple fields and presence of heterogeneities create complex dose distributions that need three dimensional dosimetry. In this work, we investigated MR-based MAGIC gel dosimetry as a three-dimensional dosimetry technique to measure the delivered dose to bladder and rectum in prostate radiation therapy. Materials and Methods: A heterogeneous slab phantom including bones was made....
متن کاملThe Effect of Breast Reconstruction Prosthesis on Photon Dose Distribution in Breast Cancer Radiotherapy
Introduction: Siliconeprosthetic implants are commonlyutilizedfor tissue replacement and breast augmentation after mastectomy. On the other hand, some patients require adjuvant radiotherapy in order to preventlocal-regional recurrence and increment ofthe overall survival. In case of recurrence, the radiation oncologist might have to irradiate the prosthesis.The aim of this study was to evaluate...
متن کاملComparison of Absorbed Dose in Thyroid and Lens as Organs at Risk Between in vivo Dosimetry and 3-D Treatment Planning Calculation in Head and Neck Radiotherapy by Linac Beam
Introduction: Critical organs and structures may receive significant amounts of irradiation even if they are not the target of radiotherapy or located outside the treatment field. Although the sensitive thyroid gland and lens are not directly the targets of treatment, they can be affected by irradiation during the treatment of tumours in head and neck region. The purpose of thi...
متن کاملEstimation of the risk for the second cancers induction in healthy tissues during the radiation therapy of liver’s tumor
The aim of this study was to estimate the delivered doses to tissues and organs around the liver tumor during radiotherapy and to determine organs with a high probability of secondary induction cancers. This study was carried out using Monte Carlo method. The used Phantom, the ORNL phantom is for an adult female. The volume of the tumor is as much as 30% of the liver tissue.In the initial step...
متن کامل