Biomechanical impact of C2 pedicle screw length in an atlantoaxial fusion construct

نویسندگان

  • Risheng Xu
  • Mohamad Bydon
  • Mohamed Macki
  • Stephen M. Belkoff
  • Evan R. Langdale
  • Kelly McGovern
  • Jean-Paul Wolinsky
  • Ziya L. Gokalsan
  • Ali Bydon
چکیده

BACKGROUND Posterior, atlantoaxial (AA) fusions of the cervical spine may include either standard (26 mm) or short (16 mm) C2 pedicle screws. This manuscript focused on an in vitro biomechanical comparison of standard versus short C2 pedicle screws to perform posterior C1-C2 AA fusions. METHODS Twelve human cadaveric spines underwent C1 lateral mass screw and standard C2 pedicle screw (n = 6) versus short C2 pedicle screw (n = 6) fixation. Six additional controls were not instrumented. The peak torque, peak rotational interval, and peak stiffness of the constructs were analyzed to failure levels. RESULTS The peak torque to construct failure was not statistically significantly different among the control spine (12.2 Nm), short pedicle fixation (15.5 Nm), or the standard pedicle fixation (11.6 Nm), P = 0.79. While the angle at the peak rotation statistically significantly differed between the control specimens (47.7° of relative motion) and the overall instrumented specimens (P < 0.001), the 20.7° of relative rotation in the short C2 pedicle screw specimens was not statistically significantly higher than the 13.7° of relative rotation in the standard C2 pedicle screw specimens (P = 0.39). Similarly, although the average stiffness was statistically significantly lower in control group (0.026 Nm/degree) versus the overall instrumented specimens (P = 0.001), the standard C2 pedicle screws (2.54 Nm/degree) did not differ from the short C2 pedicle screws. CONCLUSIONS Both standard and short C2 pedicle screws allow for equally rigid fixation of C1 lateral mass-C2 AA fusions. Usage of a short C2 pedicle screw may be an acceptable method of stabilization in carefully selected patient populations.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Posterior Atalntoaxial Fusion with C1 Lateral Mass Screw and C2 Pedicle Screw Supplemented with Miniplate Fixation for Interlaminar Fusion : A Preliminary Report

OBJECTIVE To investigate the feasibility of C1 lateral mass screw and C2 pedicle screw with polyaxial screw and rod system supplemented with miniplate for interlaminar fusion to treat various atlantoaxial instabilities. METHODS After posterior atlantoaxial fixation with lateral mass screw in the atlas and pedicle screw in the axis, we used 2 miniplates to fixate interlaminar iliac bone graft ...

متن کامل

“Direct vision” operation of posterior atlantoaxial transpedicular screw fixation for unstable atlantoaxial fractures

BACKGROUND The posterior screw fixation in atlas via posterior arch and lateral mass, also called C1 "pedicle" screw, combined with C2 pedicle screw fixiation has shown better biomechanical stability in unstable atlantoaxial fractures. However, its popularization has to fulfill the limitation imposed by anatomical characteristics. The aim of this study was to explore the manipulation, effect, a...

متن کامل

Posterior atlantal lateral mass fixation technique with polyaxial screw and rod fixation system.

OBJECTIVE Atlantoaxial instability may result from various pathologic conditions and operative treatment may be required to correct the deformity, provide stability and prevent neurological deficits. We presented our clinic's experience using C1-C2 fusion with polyaxial screw and rod fixation for C1 and C2 instability for various reasons. METHODS AND MATERIAL Four patients with atlantoaxial i...

متن کامل

Biomechanical analysis comparing three C1-C2 transarticular screw salvaging fixation techniques.

STUDY DESIGN This is an in vitro biomechanical study. OBJECTIVE To compare the biomechanical stability of the 3 C1-C2 transarticular screw salvaging fixation techniques. SUMMARY AND BACKGROUND DATA Stabilization of the atlantoaxial complex is a challenging procedure because of its complicated anatomy. Many posterior stabilization techniques of the atlantoaxial complex have been developed wi...

متن کامل

Unilateral C1 Lateral Mass and C2 Pedicle Screw Fixation for Atlantoaxial Instability in Rheumatoid Arthritis Patients: Comparison with the Bilateral Method

OBJECTIVE Bilateral C1 lateral mass and C2 pedicle screw fixation (C1LM-C2P) is an ideal technique for correcting atlantoaxial instability (AAI). However, the inevitable situation of vertebral artery injury or unfavorable bone structure may necessitate the use of unilateral C1LM-C2P. This study compares the fusion rates of the C1 lateral mass and C2 pedicle screw in the unilateral and bilateral...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 5  شماره 

صفحات  -

تاریخ انتشار 2014