Automated external cardioversion defibrillation monitoring in cardiac arrest: a randomized trial

نویسندگان

  • Bakhtiar Ali
  • Heather Bloom
  • Emir Veledar
  • Dorothy House
  • Robert Norvel
  • Samuel C Dudley
  • A Maziar Zafari
چکیده

BACKGROUND In-hospital cardiac arrest has a poor prognosis despite active electrocardiography monitoring. The initial rhythm of approximately 25% of in-hospital cardiopulmonary resuscitation (CPR) events is pulseless ventricular tachycardia/ventricular fibrillation (VT/VF). Early defibrillation is an independent predictor of survival in CPR events caused by VT/VF. The automated external cardioverter defibrillator (AECD) is a device attached by pads to the chest wall that monitors, detects, and within seconds, automatically delivers electric countershock to an appropriate tachyarrhythmia. STUDY OBJECTIVES To evaluate safety of AECD monitoring in hospitalized patients. To evaluate whether AECDs provide earlier defibrillation than hospital code teams. METHODS The study is a prospective trial randomizing patients admitted to the telemetry ward to standard CPR (code team) or standard CPR plus AECD monitoring (PowerHeart CRM). The AECD is programmed to deliver one 150 J biphasic shock to patients in sustained VT/VF. Data is collected using the Utstein criteria for cardiac arrest. The primary endpoint is time-to-defibrillation; secondary outcomes include neurological status and survival to discharge, with 3-year follow-up. RESULTS To date, 192 patients have been recruited in the time period between 10/10/2006 to 7/20/2007. A total of 3,655 hours of telemetry data have been analyzed in the AECD arm. The AECD has monitored ambulatory telemetry patients in sinus rhythm, sinus tachycardia, supraventricular tachycardia, atrial flutter or fibrillation, with premature ventricular complexes and non-sustained VT without delivery of inappropriate shocks. One patient experienced sustained VT during AECD monitoring, who was successfully defibrillated (17 seconds after meeting programmed criteria). There are no events to report in the control arm. The patient survived the event without neurological complications. During the same time period, mean time to shock for VT/VF cardiac arrest occurring outside the telemetry ward was 230 +/- 50 seconds. CONCLUSION AECD monitoring is safe and likely results in earlier defibrillation than standard telemetry monitoring. TRIAL REGISTRATION National Institutes of Health registration ID: NCT00382928.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

BIPHASIC Trial: a randomized comparison of fixed lower versus escalating higher energy levels for defibrillation in out-of-hospital cardiac arrest.

BACKGROUND There is little clear evidence as to the optimal energy levels for initial and subsequent shocks in biphasic waveform defibrillation. The present study compared fixed lower- and escalating higher-energy regimens for out-of-hospital cardiac arrest. METHODS AND RESULTS The Randomized Controlled Trial to Compare Fixed Versus Escalating Energy Regimens for Biphasic Waveform Defibrillat...

متن کامل

A Randomized Comparison of Fixed Lower Versus Escalating Higher Energy Levels for Defibrillation in Out-of-Hospital Cardiac Arrest

Background—There is little clear evidence as to the optimal energy levels for initial and subsequent shocks in biphasic waveform defibrillation. The present study compared fixed lowerand escalating higher-energy regimens for out-of-hospital cardiac arrest. Methods and Results—The Randomized Controlled Trial to Compare Fixed Versus Escalating Energy Regimens for Biphasic Waveform Defibrillation ...

متن کامل

Automated External Defibrillators, Defibrillation, Cardioversion, and Pacing 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care

Defibrillation Plus CPR: A Critical Combination Early defibrillation is critical to survival from sudden cardiac arrest (SCA) for several reasons1: the most frequent initial rhythm in out-of-hospital witnessed SCA is ventricular fibrillation (VF),2 the treatment for ventricular fibrillation is defibrillation,3 the probability of successful defibrillation diminishes rapidly over time,4 and VF te...

متن کامل

The Role of Automated External Defibrillators in Athletics

CONTEXT Sudden cardiac arrest is the leading cause of death in young athletes. The purpose of this review is to summarize the role of automated external defibrillators and emergency planning for sudden cardiac arrest in the athletic setting. EVIDENCE ACQUISITION Relevant studies on automated external defibrillators, early defibrillation, and public-access defibrillation programs were reviewed...

متن کامل

Part 6: electrical therapies: automated external defibrillators, defibrillation, cardioversion, and pacing: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

The recommendations for electrical therapies described in this section are designed to improve survival from SCA and life-threatening arrhythmias. Whenever defibrillation is attempted, rescuers must coordinate high-quality CPR with defibrillation to minimize interruptions in chest compressions and to ensure immediate resumption of chest compressions after shock delivery. The high first-shock ef...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Trials

دوره 9  شماره 

صفحات  -