Modelling the Interaction Forces between an Ideal Measurement Surface and the Stylus Tip of a Novel Vibrating Micro-scale CMM Probe

نویسندگان

  • James D. Claverley
  • A. Georgi
  • Richard K. Leach
چکیده

This paper describes the development of an analytical model to describe a novel three-axis vibrating micro-scale probe for micro-coordinate measuring machines (micro-CMMs). The micro-CMM probe is vibrated in three axes, in order to address the problems inherent with microand nano-scale co-ordinate measurements caused by surface interaction forces. These surface forces have been investigated and a mathematical model describing an ideal probing situation has been developed. The vibration amplitude required for the probe to overcome the surface interaction forces has been calculated using this model. The results of initial vibration experiments are reported and the suitability of the probe to counteract the surface interaction forces is confirmed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dynamic modeling and nonlinear vibration simulation of piezoelectric micro-beam in self sensing mode of atomic force microscope

Nowadays, atomic force microscope is considered as a useful tool in the determination of intermolecular forces and surface topography with the resolution of nanometers. In this kind of microscope, micro cantilever is considered as the heart of the microscope and is used as a measuring tool.  This paper is aimed towards investigating the behavior of a piezoelectric micro cantilever with a triang...

متن کامل

Performance Assessment of a New Variable Stiffness Probing System for Micro-CMMs

When designing micro-scale tactile probes, a design trade-off must be made between the stiffness and flexibility of the probing element. The probe must be flexible enough to ensure sensitive parts are not damaged during contact, but it must be stiff enough to overcome attractive surface forces, ensure it is not excessively fragile, easily damaged or sensitive to inertial loads. To address the n...

متن کامل

Probing System Characteristics in Coordinate Metrology

This paper aims at studying the effect of the dynamic errors on surface measurements using three different types of touch trigger probes attached to a bridge-type-CMM. Unforeseeable dynamic root errors of a ductile touch trigger probing system have been characterized theoretically and experimentally as well. The results were employed in validating a developed analytical twodimensional-model (2D...

متن کامل

Accessing Time-Varying Forces on the Vibrating Tip of the Dynamic Atomic Force Microscope to Map Material Composition

In dynamic atomic force microscopes the primary physical quantities being measured are the amplitude/phase or amplitude/frequency of the vibrating force probe. Topographic images with spatial resolutions down to the atomic scale can be obtained by mapping these measurements across the sample surface under feedback control. During the imaging process the vibrating tip is observing tip–sample int...

متن کامل

A scanning contact probe for micro CMM

A new high precision contact scanning probe possible to measure miniature components on a micro/nano-CMM is proposed. This contact probe is composed of a fiber stylus with a ball tip, a floating plate and focus sensors. The ball tip is fabricated using optical fiber with melting and solidification processes. The stylus is attached to a floating plate, which is connected to the probe housing via...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2010