Powered bone marrow biopsy procedures produce larger core specimens, with less pain, in less time than with standard manual devices
نویسندگان
چکیده
Bone marrow sampling remains essential in the evaluation of hematopoietic and many non-hematopoietic disorders. One common limitation to these procedures is the discomfort experienced by patients. To address whether a Powered biopsy system could reduce discomfort while providing equivalent or better results, we performed a randomized trial in adult volunteers. Twenty-six subjects underwent bilateral biopsies with each device. Core samples were obtained in 66.7% of Manual insertions; 100% of Powered insertions (P=0.002). Initial mean biopsy core lengths were 11.1±4.5 mm for the Manual device; 17.0±6.8 mm for the Powered device (P<0.005). Pathology assessment for the Manual device showed a mean length of 6.1±5.6 mm, width of 1.0±0.7 mm, and volume of 11.0±10.8 mm(3). Powered device measurements were mean length of 15.3±6.1 mm, width of 2.0±0.3 mm, and volume of 49.1±21.5 mm(3) (P<0.001). The mean time to core ejection was 86 seconds for Manual device; 47 seconds for the Powered device (P<0.001). The mean second look overall pain score was 33.3 for the Manual device; 20.9 for the Powered (P=0.039). We conclude that the Powered biopsy device produces superior sized specimens, with less overall pain, in less time.
منابع مشابه
Using a Powered Bone Marrow Biopsy System Results in Shorter Procedures, Causes Less Residual Pain to Adult Patients, and Yields Larger Specimens
BACKGROUND In recent years, a battery-powered bone marrow biopsy system was developed and cleared by the U.S. Food and Drug Administration to allow health care providers to access the bone marrow space quickly and efficiently. A multicenter randomized clinical trial was designed for adult patients to determine if the powered device had advantages over traditional manually-inserted needles in re...
متن کاملComparison of a powered bone marrow biopsy device with a manual system: results of a prospective randomised controlled trial
The diagnostic and clinical usefulness of a powered bone marrow biopsy device (OnControl()) versus a standard manual device (TRAP Hospital System) was studied. Primary endpoints were biopsy quality and patient pain during the procedure. Fifty patients underwent a total of 60 procedures by three expert operators in a randomised stratified fashion. Baseline demographic and clinical parameters wer...
متن کاملA powered bone marrow biopsy system versus manual methods: a systematic review and meta-analysis of randomised trials
A significant number of bone marrow biopsies are performed annually. Outcomes of patient pain and sample size of biopsy are important issues as perception of patient pain is underestimated by clinicians and sample size assists in accurate diagnosis. Manual extraction of marrow biopsy is the main method used. Recently a powered system has been introduced. The objectives of this systematic review...
متن کاملThe OnControl bone marrow biopsy technique is superior to the standard manual technique for hematologists-in-training: a prospective, randomized comparison
The purpose of this study was to compare a novel bone marrow device with the standard marrow needle in a prospective, randomized study in a teaching hospital employing hematologists-in-training. The new device, the OnControl Bone Marrow (OBM) Biopsy System, utilizes a battery-powered drill to insert the needle. Fifty-four bone marrows (27 standard and 27 OBM) were performed by 11 fellows under ...
متن کاملGeneral Anesthesia for Lumbar Puncture and Bone Marrow Aspiration /Biopsy in Children
Background Multiple procedures (Lumbar puncture and bone marrow aspiration /biopsy) cause pain, stress, depression and etc for the patients and their families. Various methods have been recommended for pain reduction during invasive procedures. The aim of this study is to report the complications following general anesthesia. Methods In this prospective observational study, two hundred and ...
متن کامل