Electrical impedance tomography: Amplitudes of cardiac related impedance changes in the lung are highly position dependent
نویسندگان
چکیده
BACKGROUND Electrical impedance tomography (EIT) is used on the thorax to measure impedance changes due to the presence of air and blood in the lung. This experimental study was performed to investigate the effect of posture on cardiac and respiratory related impedance changes. METHODS EIT measurements were performed on 14 healthy subjects in left-, right lateral, prone, supine and upright positions. Simultaneously, tidal volume was recorded with an ultrasonic flowmeter. For image reconstruction, the classic Sheffield back-projection and three variants of the modern GREIT algorithm were applied with two different reference frames. Amplitudes of cardiac- and respiratory impedance changes were extracted and compared between the positions. RESULTS We found significant differences in both cardiac and respiratory amplitudes between postures. Especially, supine and upright positions showed dramatic changes in amplitude. These differences between postures were unaffected by the change of reference frames in all reconstruction methods except of the classic Sheffield back projection. Possible sources that explain the observed posture dependency are discussed. CONCLUSION Researchers and clinicians need to be aware of this phenomenon when comparing EIT amplitudes in different body positions.
منابع مشابه
Cardiac-related impedance changes obtained by electrical impedance tomography: an acceptable parameter for assessment of pulmonary perfusion?
monitoring regional ventilation. Interest is growing to derive additional information on pulmonary perfusion and ventilation/perfusion distribution. Since signals recorded by electrical impedance tomography also contain cardiac-related impedance changes, attempts are made to evaluate them in view of perfusion. Recently in Critical Care a corresponding study applying an advanced fi ltering techn...
متن کاملApplications of Electrical Impedance Tomography in Neurology
Introduction: Electrical impedance tomography (EIT) is a non-invasive technique utilized in various medical applications, including brain imaging and other neurological diseases. Recognizing the physiological and anatomical characteristics of organs based on their electrical properties is one of the main applications of EIT, as each variety of tissue structure has its own electrical characteris...
متن کاملEffect of Electrode Belt and Body Positions on Regional Pulmonary Ventilation- and Perfusion-Related Impedance Changes Measured by Electric Impedance Tomography
Ventilator-induced or ventilator-associated lung injury (VILI/VALI) is common and there is an increasing demand for a tool that can optimize ventilator settings. Electrical impedance tomography (EIT) can detect changes in impedance caused by pulmonary ventilation and perfusion, but the effect of changes in the position of the body and in the placing of the electrode belt on the impedance signal...
متن کاملEffect of high-flow nasal cannula and body position on end-expiratory lung volume: a cohort study using electrical impedance tomography.
BACKGROUND Electrical impedance tomography measures changes in lung impedance, which are mainly related to changes in lung volume. We used electrical impedance tomography to investigate the effects of high-flow nasal cannula (HFNC) and body position on global and regional end-expiratory lung impedance variation (ΔEELI). METHODS Prospective study with 20 healthy adults. Two periods were define...
متن کاملAccuracy and reliability of noninvasive stroke volume monitoring via ECG-gated 3D electrical impedance tomography in healthy volunteers
Cardiac output (CO) and stroke volume (SV) are parameters of key clinical interest. Many techniques exist to measure CO and SV, but are either invasive or insufficiently accurate in clinical settings. Electrical impedance tomography (EIT) has been suggested as a noninvasive measure of SV, but inconsistent results have been reported. Our goal is to determine the accuracy and reliability of EIT-b...
متن کامل