GFR Determination by a Modification of the Gates Method: The Conventional Renal Examination with a Semi-Automated GFR Measurement
نویسندگان
چکیده
The conventional radionuclide renal examination using a 1020-mCi dose of technetium-99m t mTc) DTPA (Pentetate), conducted over a 30-60-min duration, does not accommodate the Gates method for GFR determination. Simple modifications to the Gates technique will, however, permit a GFR determination to be made without sacrificing the components of a complete conventional renal examination, and will not significantly alter the technologist's examination routine or time commitment. These modifications allow the Gates method to be used on virtually any gamma camera/collimator combination with an accompanying computer system, dose calibrator, and a lead slab syringe attenuator. With the modifications, GFR determinations can be obtained using doses of tracer as high as 10-20 mCi, while eliminating the direct gamma camera counting of the residual activity in the postinjection syringe. The net injected syringe counts are calculated using the gamma camera counts from the preinjection syringe attenuated through lead and the dose calibrator's preand postinjection measurements of activity to determine the fraction of injected counts from the measured syringe dose. Excellent results are obtained when the gamma camera counts of the syringe activity using the original Gates method are compared to the counts derived from the modified technique using the lead slab attenuator and the dose calibrator (r > 0.99). We have found that this particular modification of the Gates method allows for a full examination by removing the dose and time restrictions imposed by the original Gates technique.
منابع مشابه
Optimization of Gate’s technique for measurement of GFR during routine renal scan [Persian]
Using Gate’s technique for GFR measurement with Gamma camera, the patient is received 5mCi of Tc-99m-DTPA. By this amount of radioactivity, good quality renal scan is not possible. This study tries to optimize Gate’s technique for GFR measurement during routine renal scan (with 10-15 mCi). Methods and materials: Scanning was performed for 60 seconds from samples of Tc-99m with activities ...
متن کاملDetermining an accurate method to estimate GFR in renal transplant recipients with stable serum creatinine levels
Introduction:Detecting renal allograft dysfunction early will allow timely diagnosis and treatment. There is no objective recommendation by national kidney societies for glomerular filtration rate (eGFR) estimation in post-transplant setting. 99mTc-DTPA Technetium-99m Diethylene triamine penta acetic acid) renogram can identify early renal dysfunction much before ser...
متن کاملQuantitative and Qualitative Scintigraphic Measurement of Renal Function in Persian Cat Using 99mTechnetium Ethylenedicysteine and 99mTechnetium Diethylenetriaminepentaacetic
Objective- To consider the use of renal scintigraphic evaluation of kidney function in Persian cat and make a comparison between 99mTechnetium ethylenedicysteine and 99mTechnetium diethylenetriaminepentaacetic in renal scintigraphy. Design- Descriptive study Animals- 6 male healthy adult Persian cats - no clinical sign of renal disorders and prior to presentation. Proce...
متن کاملAccuracy of modification of diet for renal disease and Cockcroft-Gault equations as compared to the radioisotope double sample method: A study in patients with acute renal failure
Introduction: We intended to assess the accuracy of re-expressed Modification of Diet for Renal Disease (MDRD) and Cockcroft-Gault (CG) equations to estimate glomerular filtration rate (GFR) in chronic kidney disease in two different etiologies of acute renal failure (ARF): acute tubular necrosis (ATN) and acute glomerulonephritis (AGN). Methods: Patients admitted for ARF or the patients compl...
متن کاملGlomerular filtration rate determination by creatinine and cystatin-C in patients with acute pyelonephritis
Background: Measurement of glomerular filtration rate (GFR) and monitoring of it in any patient on nephrotoxic drugs is very important. Recently, cystatin C (cys-C) has been introduced as a better marker for determining and monitoring renal function than creatinine especially in a mild decrease of GFR. This study was done to assess the change of GFR measurement based on serum Cys-C and creatini...
متن کامل