Performance of Transport Ventilators
ثبت نشده
چکیده
The paper by Chipman et al in the June 2007 issue of RESPIRATORY CARE represented a monumental undertaking to evaluate 15 transport ventilators.1 We reviewed the paper on several occasions to reconcile their findings with our work on transport ventilators.2-9 We write to point out several inaccuracies in the study and ask Chipman et al to more clearly explain their findings and recommendations. In their Table 1, Chipman et al list the positive end-expiratory pressure (PEEP) of the LTV1000 as 0–30 cm H2O. The actual range is 0–20 cm H2O. The table also lists the PEEP of the Crossvent 3 as 0 cm H2O, but the Crossvent 3 is capable of PEEP up to 35 cm H2O. Were Chipman et al referring to an earlier model of the Crossvent 3 that couldn’t supply PEEP? Also, the Vortran is listed as being capable of providing continuous mandatory ventilation (CMV) and intermittent mandatory ventilation (IMV), but in fact the Vortran can provide only pressure-cycled CMV. Similarly, the Percussionaire TXP provides only IMV, not CMV. With the Percussionaire TXP, if the patient breathes spontaneously, air is entrained from the inspiratory port of the Phasitron. We are unclear as to why Chipman et al listed both assist-control and CMV for some ventilators, as those 2 modes are the same. Did they mean to distinguish between ventilators that can be patient-triggered and those that can deliver only mandatory breaths? If so, then errors still remain. The Pneupac ParaPac ventilators have a mode termed SMMV (synchronized minimum mandatory volume), not MMV (mandatory minute volume) or IMV, which are quite different, in that if the patient breathes spontaneously at a respiratory rate equal to or greater than the set rate, the ventilator does not provide any mandatory breaths. Also the Crossvent 3 has a selection for pressure support but does not have the ability to flow cycle. This mode results in patient-triggered, pressure-limited, timecycled support. The Crossvent 3 is listed as having both volume and pressure control, but in fact it is capable of only pressurelimited ventilation via a mechanical pressure-relief valve. In that instance the ventilator continues to deliver the set flow and volume according to the volume control settings, but vents gas to the atmosphere if the pressure threshold is reached. These errors may have resulted from Chipman et al depending on the manufacturer’s literature for their data. However, since their intention was to compare available devices, we believe these errors should be pointed out so readers can determine the actual performance of the devices, not simply mimic the manufacturer’s specifications. Chipman et al stated that in their bench protocol the Impact/Uni-Vent Eagle 754 operated on an E-size cylinder for 35 min, which is substantially less than we have observed and is not consistent with that ventilator’s function.2 Was this test repeated more than once? This finding suggests either a leak in the system, a malfunctioning ventilator, or unfamiliarity by the operator. Was the tidal volume (VT) continuously measured to assure that the correct minute volume was being delivered? As Chipman et al are well aware from their experience, repetition of laboratory experiments is invaluable in the detection of one-time errors. In Table 2, Chipman et al concluded that neither the Impact/Uni-Vent Eagle 754 nor thePulmoneticLTV1000arecapableofventilating injured lungs. As in our previous comment, that conclusion is inconsistent with the bench data and the worldwide experience with those 2 devices. For over a decade the Impact/Uni-Vent Eagle 754 has been successfully used to transport critically ill soldiers (including many with acute respiratory distress syndrome) between military care centers. Similarly, the LTV1000 is capable of volume and pressure control at up to 80 breaths/min, and is used by many centers for transport of critically ill patients. We request that Chipman et al explain the nature of the failures with the Impact/UniVent Eagle 754 and the Pulmonetic LTV1000, and reconcile their findings with the common experience. Our review of the animal-model experiments in the Chipman et al study led us to understand that a single animal was used to test 5 separate ventilators, in random sequence. Is it possible that the small number of animals resulted in a ventilator being disadvantaged in the trial because of the sequence? As an example, Figure 2 indicates that each ventilator was used in 2 animal experiments. What was the progression of the lung injury? If the first 4 ventilators failed to provide adequate oxygenation or ventilation, would that worsen the lung injury and thus make the fifth ventilator fail to provide adequate gas exchange? Does their surfactant-wash-out model improve with time, or does gas exchange steadily worsen? Also, the performance of every ventilator is affected by the device characteristics, the condition of the animal, and the understanding of the operator. It seems incongruent that a given ventilator is capable of maintaining gas exchange in one animal, but not the other, unless the ventilator model is substantially different or the operator fails to make the appropriate changes. Finally, though Chipman et al concluded that the Impact/Uni-Vent Eagle 754 is unable to ventilate injured lungs, it is one of the ventilators they recommend in their conclusions. Though we agree, and experience clearly demonstrates that this recommendation makes sense, how did they come to this conclusion? At the end of their exhaustive study, how did Chipman et al determine which ventilators to recommend? Were characteristics ranked or weighted? Was eachventilatorgivenascore forperformance in each of the evaluations? Though we agree that both the Impact/Uni-Vent Eagle 754 and Newport HT50 meet the requirements for a “front-line ventilator for rescue situations,” we are not sure why the LTV1000 and VersaMed iVent do not.10 Did Chipman et al eliminate the LTV1000 because of its high gas consumption? Was the VersaMed eliminated because of short battery life and excessive weight? We request that Chipman et al describe the system they used to come to their conclusions. If gas consumption, size, and battery life were the only factors considered, the animal experiments seem unnecessary. We appreciate the substantial effort by Chipman et al in this project and their commitment to provide the respiratory care community with much-needed data. However, we believe they should correct some errors and explain the findings that seem to run
منابع مشابه
A comparison of turbine-driven mechanical ventilators and conventional ventilators
Turbine-driven mechanical ventilators have some important advantages over conventional ventilators. The most obvious of these is independence of compressed air supply during patient transport, but higher NIV performance due to higher peak flow is also often mentioned. However, there is still a widespread belief that conventional ventilators using compressed air supplies outperform turbine-drive...
متن کاملEvaluation of ventilators used during transport of critically ill patients: a bench study.
OBJECTIVE To evaluate the most recent transport ventilators' operational performance regarding volume delivery in controlled mode, trigger function, and the quality of pressurization in pressure support mode. METHODS Eight recent transport ventilators were included in a bench study in order to evaluate their accuracy to deliver a set tidal volume under normal resistance and compliance conditi...
متن کاملPerformance comparison of 15 transport ventilators.
BACKGROUND Numerous mechanical ventilators are designed and marketed for use in patient transport. The complexity of these ventilators differs considerably, but very few data exist to compare their operational capabilities. METHODS Using bench and animal models, we studied 15 currently available transport ventilators with regard to their physical characteristics, gas consumption (duration of ...
متن کاملPerformance of transport ventilator with patient-triggered ventilation.
OBJECTIVES Transport ventilators with inspiratory triggering functions and pressure support-control modes have recently become commercially available. We evaluated these ventilators in comparison with a standard ICU ventilator. STUDY DESIGN Laboratory study with a mechanical lung model. METHODS We compared the performance of four transport ventilators (model 740, Mallinckrodt, Pleasanton, C...
متن کاملBench-test comparison of 26 emergency and transport ventilators
INTRODUCTION Numerous emergency and transport ventilators are commercialized and new generations arise constantly. The aim of this study was to evaluate a large panel of ventilators to allow clinicians to choose a device, taking into account their specificities of use. METHODS This experimental bench-test took into account general characteristics and technical performances. Performances were ...
متن کاملThe performance of Dräger Oxylog ventilators at simulated altitude.
Ventilated patients frequently require transport by air in a hypobaric environment. Previous studies have demonstrated significant changes in the performance of ventilators with changes in cabin pressure (altitude) but no studies have been published on the function of modem ventilators at altitude. This experiment set out to evaluate ventilatory parameters (tidal volume and respiratory rate) of...
متن کامل