Lithium determinations evaluated in eight analyzers.

نویسندگان

  • M Sampson
  • M Ruddel
  • R J Elin
چکیده

We compared five ion-selective electrodes (ISEs; AVL 985S, Baxter Lytening 2Z, Beckman EL-ISE, Ciba-Corning 654 Na/K/Li, and Nova CRT 11) and a colorimetric method (Ektachem) for determination of lithium with flame atomic absorption and atomic emission spectroscopy. We evaluated precision, recovery, interference by drugs (procainamide, N-acetylprocainamide, quinidine, lidocaine, carbamazepine, and valproic acid) and inorganic analytes (Na, K, Ca, Mg, and Br), and performance with sera from patients receiving lithium. Imprecision was < 5% (CV) for all analyzers except Ektachem and Beckman. Analytical recovery was 100% +/- 5% for all analyzers except Ektachem and Baxter. Some drug interference was seen with all analyzers except AVL and Corning. Calcium caused interference with AVL, Corning, and Ektachem analyzers, and sodium and potassium interfered with the Ektachem analyzer. The results with the Baxter, Beckman, and Corning analyzers were closest to those by flame atomic emission (mean differences, 0.01 to 0.02 mmol/L); the AVL and Nova analyzers gave lower results (mean differences, -0.11 to -0.13 mmol/L) and the Ektachem gave higher results (mean difference, 0.08 mmol/L).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Spherical analyzers and monochromators for resonant inelastic hard X-ray scattering: a compilation of crystals and reflections

Resonant inelastic X-ray scattering (RIXS) experiments require special sets of near-backscattering spherical diced analyzers and high-resolution monochromators for every distinct absorption-edge energy and emission line. For the purpose of aiding the design and planning of efficient RIXS experiments, comprehensive lists of suitable analyzer reflections for silicon, germanium, α-quartz, sapphire...

متن کامل

Peykan check, a simple continuous quality control method for hematology analyzers

A simple inexpensive con!nuous quality control method, by means of eight basic blood coun!ng parameters, obtained from automated hematology analyzers, using pa!ent samples, is described. A few samples with low, normal and high values were selected and introduced to the instrument early in the morning as the first count, the results of which are plo#ed on the appropriate chart as dots. The sa...

متن کامل

Evaluation of the pocH-100iV DIFF hematology analyzer for use in horses and cattle Evaluatie van de pocH-100iV DIFF-hematologie analyzer voor paarden- en runderbloed

The results of the analysis of equine and bovine blood samples with the automated pocH-100iV DIFF hematology analyzer were compared with the results obtained with reference methods or other analyzers (Vet ABC, Coulter Counter ZF, Coulter LH 750 analyzer). For equine blood and most parameters in bovine blood good to excellent correlations between methods and analyzers were obtained. For bovine b...

متن کامل

اثر توپیرامات بر کنترل وزن مبتلایان به اختلال دوقطبی نوع یک تحت درمان با لیتیوم و داروهای ضد روان‌پریشی در دوره شیدایی

  Objectives : The purpose of this study was to evaluate the role of topiramate in weight gain control in bipolar patients receiving lithium carbonate and antipsychotics during their manic episode. Method: In current randomized controlled double-blind clinical trial, 46 patients with bipolar disorder in their manic episode were divided into two intervention and control groups. Both groups were ...

متن کامل

Inter-instrument comparison of blood gas analyzers and assessment of tonometry using fresh heparinized whole human blood.

Because of limitations associated with commercial blood gas quality controls and tonometry of stored blood, fresh heparinized blood was used to compare PO2 and PCO2 performance of ten blood gas analyzers. Function of nine gas mixer/tonometer systems was evaluated. These were used to create blood samples with target values for PO2 and PCO2. All ten analyzers had high precision; this magnified sm...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 40 6  شماره 

صفحات  -

تاریخ انتشار 1994