Optimization of Saddle Coils for Magnetic Resonance Imaging
نویسندگان
چکیده
In Nuclear Magnetic Resonance (NMR) experiments, besides the apparatus designed to acquire the NMR signal, it is necessary to generate a radio frequency electromagnetic field using a device capable to transduce electromagnetic power into a transverse magnetic field. We must generate this transverse homogeneous magnetic field inside the region of interest with minimum power consumption. Many configurations have been proposed for this task, from coils to resonators. For low field intensity (<0.5 T) and small sample dimensions (<30 cm), the saddle coil configuration has been widely used. In this work we present a simplified method for calculating the magnetic field distribution in these coils considering the current density profile. We propose an optimized saddle configuration as a function of the dimensions of the region of interest, taking into account the uniformity and the sensitivity. In order to evaluate the magnetic field uniformity three quantities have been analyzed: Non-uniformity, peak-to-peak homogeneity and relative uniformity. Some experimental results are presented to validate our calculation.
منابع مشابه
Non-uniformity of Clinical Head, Head and Neck, and Body Coils in Magnetic Resonance Imaging (MRI)
Introduction Signal intensity uniformity in a magnetic resonance (MR) image indicates how well the MR imaging (MRI) system represents an object. One of the major sources of image non-uniformity in high-field MRI scanners is inhomogeneity of radio-frequency coil. The aim of this study was to investigate non-uniformity in head, head and neck, and body coils and compare the obtained results to det...
متن کاملTransceiver double crossed saddle for rodents at 2T
Introduction: Small animal MR imaging and spectroscopy, where spatial resolution is a limiting factor, requires the use of distinct transmit and receive coils [1]. This approach makes easier the optimization of the RF field homogeneity for transmit-only coils and signal-to-noise ratio (SNR) for receive-only coils. It also requires that RF coils should be electrically decoupled, commonly using P...
متن کاملDesigning and Fabrication of a New Radiofrequency Planar microcoil for mini-Nuclear Magnetic Resonance
Introduction Radiofrequency planar microcoils are used to increase the resolution of magnetic resonance images of small samples. In this study, we aimed to design and fabricate a spiral planar microcoil constructed on a double-sided printed circuit board (PCB). It has four rings with an internal diameter of 241 microns tuned and matched at 63.8 MHz. Materials and Methods To achieve the maximum ...
متن کاملPhase and sensitivity of receiver coils in magnetic resonance imaging.
Receiver coil response is a major cause of nonuniformities in magnetic resonance images. The spatial dependence of the sensitivity and phase of single-saddle receiver coils has been investigated quantitatively by calculating the H1 field and comparing the results with measurements of a uniform phantom. Agreement between the measurements and calculations is excellent. A method is developed which...
متن کاملMagnetic Field of RF Coil Measurement by NMR Method
A spectroscopic imaging technique with high spatial resolution was used for the measurement of the static magnetic field homogeneity of RF coils. Measured magnetic fields of optimised and not optimised saddle-shaped coils were compared with the theoretical calculations. Results are in good qualitative agreement between calculated and measured magnetic field distributions. Quantitative disagreem...
متن کامل