Insertion torque is not a good predictor of pedicle screw loosening after spinal instrumentation: a prospective study in 8 patients
نویسندگان
چکیده
BACKGROUND Pedicle screw loosening is a major safety concern in instrumented spinal surgery due to loosening with potential pseudarthrosis and possible loss of correction requiring revision surgery. Several cadaver studies have compared insertion torque of pedicle screws with resistance to pullout or cyclic loading. In most of these studies, a correlation has been found between these variables. Clinical studies have been made, comparing insertion torque to bone mineral density or radiological signs of screw loosening. There are no clinical studies comparing insertion torque to extraction torque or other biomechanical parameters in vivo. This study was designed to investigate whether the insertion torque of pedicle screws can be used to predict the purchase of the screws. METHODS The insertion torque of stainless steel pedicle screws was recorded in eight patients undergoing lumbar fusion surgery with four-screw constructs. Torque gauge manometers were used for the recordings. The implants were removed after one year, and the extraction torque of the screws was recorded. RESULTS The mean insertion torque was 76 ± 41 Ncm and the mean extraction torque 29 ± 36 Ncm. The r value was 0.591, suggesting that there was a correlation between the insertion and extraction torque. However, the scattergram revealed that the screws could be divided into two groups, six screws with a high correlation between insertion and extraction torque, and 26 screws where no correlation could be demonstrated. CONCLUSIONS In this unique human in-vivo study, the insertion torque could not be used to predict the purchase of lumbar pedicle screws one year after implantation. It could be demonstrated that in vivo insertion torque alone is of minor value to estimate pullout strength, and should be combined with or replaced by more accurate measures.
منابع مشابه
Intraoperative biomechanics of lumbar pedicle screw loosening following successful arthrodesis.
Pedicle screw loosening has been implicated in recurrent back pain after lumbar spinal fusion, but the degree of loosening has not been systematically quantified in patients. Instrumentation removal is an option for patients with successful arthrodesis, but remains controversial. Here, we quantified pedicle screw loosening by measuring screw insertion and/or removal torque at high statistical p...
متن کاملThe correlation between insertion torque of pedicle screws with bone mineral density values in posterior lumbar pedicle screw fixation
INTRODUCTION: In a pedicle screw fixation, the degree of osseointegration between a screw and the bone is an essential factor for the stability and fusion. In patients with osteoporosis, however, there is a high possibility for a pedicle screw loosening. Owing to this, there is a difficulty in determining the surgical indications. Previous biomechanical studies have reported that a pedicle scre...
متن کاملشکست وسایل تثبیت ستون فقرات
This was a retrospective review of a consecutive series of patients with spinal disease in year 2000, who underwent posterior fusion and instrumentation with Harrington distraction and Cotrel-Dobousset system to evaluate causes of hardware failure. Many cases of clinical failure has been observed in spinal instrumentation used in spinal disorder like spondylolisthesis, fractures, deformities, …...
متن کاملThe significance of radiolucent zones surrounding pedicle screws. Definition of screw loosening in spinal instrumentation.
We examined the radiographs from a prospective clinical study of fixation by pedicle screws and those from an experimental study in a sheep model. In the clinical study, instruments were removed from 21 patients after implantation for 11 to 16 months and the extraction torques of the screws were recorded. A structured protocol was used for the radiological examinations. In the experimental stud...
متن کاملPedicle screw surface coatings improve fixation in nonfusion spinal constructs.
STUDY DESIGN Biomechanical and histologic analysis. OBJECTIVE To compare the strength of the bone-screw interface of standard uncoated pedicle screws with screws treated with hydroxyapatite (HA), titanium plasma spray (TPS), and a composite HA-TPS coating. SUMMARY OF BACKGROUND DATA Transpedicular screw fixation has become the gold standard in the treatment of various thoracolumbar spinal c...
متن کامل