Mis-sizing of stent promotes intimal hyperplasia: impact of endothelial shear and intramural stress.

نویسندگان

  • Henry Y Chen
  • Anjan K Sinha
  • Jenny S Choy
  • Hai Zheng
  • Michael Sturek
  • Brian Bigelow
  • Deepak L Bhatt
  • Ghassan S Kassab
چکیده

Stent can cause flow disturbances on the endothelium and compliance mismatch and increased stress on the vessel wall. These effects can cause low wall shear stress (WSS), high wall shear stress gradient (WSSG), oscillatory shear index (OSI), and circumferential wall stress (CWS), which may promote neointimal hyperplasia (IH). The hypothesis is that stent-induced abnormal fluid and solid mechanics contribute to IH. To vary the range of WSS, WSSG, OSI, and CWS, we intentionally mismatched the size of stents to that of the vessel lumen. Stents were implanted in coronary arteries of 10 swine. Intravascular ultrasound (IVUS) was used to size the coronary arteries and stents. After 4 wk of stent implantation, IVUS was performed again to determine the extent of IH. In conjunction, computational models of actual stents, the artery, and non-Newtonian blood were created in a computer simulation to yield the distribution of WSS, WSSG, OSI, and CWS in the stented vessel wall. An inverse relation (R(2) = 0.59, P < 0.005) between WSS and IH was found based on a linear regression analysis. Linear relations between WSSG, OSI, and IH were observed (R(2) = 0.48 and 0.50, respectively, P < 0.005). A linear relation (R(2) = 0.58, P < 0.005) between CWS and IH was also found. More statistically significant linear relations between the ratio of CWS to WSS (CWS/WSS), the products CWS × WSSG and CWS × OSI, and IH were observed (R(2) = 0.67, 0.54, and 0.56, respectively, P < 0.005), suggesting that both fluid and solid mechanics influence the extent of IH. Stents create endothelial flow disturbances and intramural wall stress concentrations, which correlate with the extent of IH formation, and these effects were exaggerated with mismatch of stent/vessel size. These findings reveal the importance of reliable vessel and stent sizing to improve the mechanics on the vessel wall and minimize IH.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Impact of stent mis-sizing and mis-positioning on coronary fluid wall shear and intramural stress.

Stent deployments with geographical miss (GM) are associated with increased risk of target-vessel revascularization and periprocedural myocardial infarction. The aim of the current study was to investigate the underlying biomechanical mechanisms for adverse events with GM. The hypothesis is that stent deployment with GM [longitudinal GM, or LGM (i.e., stent not centered on the lesion); or radia...

متن کامل

Effects of stent sizing on endothelial and vessel wall stress: potential mechanisms for in-stent restenosis.

Stent sizing and apposition have been shown to be important determinants of clinical outcome. This study evaluates the mechanical effects of undersizing and oversizing of stents on endothelial wall shear stress (WSS) and vessel wall stress to determine a possible biomechanical mechanism of in-stent restenosis and thrombosis. Three-dimensional computational models of stents, artery, and internal...

متن کامل

Restenosis after coronary angioplasty. Potential biologic determinants and role of intimal hyperplasia.

Restenosis after successful PTCA remains a major problem limiting the efficacy of the procedure. The pathophysiologic mechanism of restenosis has been enigmatic so far, but accumulated evidence strongly suggests that intimal hyperplasia is the major mechanism. Based on current understanding of the process of intimal hyperplasia, one unifying concept may be that there are at least two major loca...

متن کامل

Wall shear stress in the development of in-stent restenosis revisited. A critical review of clinical data on shear stress after intracoronary stent implantation.

The average wall shear stress (WSS) is in 1 Pa range in coronary arteries, while the stretching effect of an implanted coronary stent can generate up to 3 × 105 times higher circumferential stress in the vessel wall. It is widely accepted that WSS plays a critical role in the development of restenosis after coronary stent implantation, but relevant clinical endpoint studies are lack-ing. Fluid ...

متن کامل

Enhanced external counterpulsation inhibits intimal hyperplasia by modifying shear stress responsive gene expression in hypercholesterolemic pigs.

BACKGROUND Enhanced external counterpulsation (EECP) is a circulation assist device that may improve endothelial dysfunction by increasing shear stress. Chronic exposure of vascular endothelial cells and vascular smooth muscle cells to relatively high physiological shear stress has antiproliferative and vasoprotective effects. The present study hypothesizes that EECP inhibits intimal hyperplasi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • American journal of physiology. Heart and circulatory physiology

دوره 301 6  شماره 

صفحات  -

تاریخ انتشار 2011