Contralateral pneumothorax after cardiac pacemaker implantation

نویسندگان

  • Małgorzata Hardzina
  • Andrzej Ząbek
  • Krzysztof Boczar
  • Paweł Matusik
  • Barbara Małecka
  • Jacek Lelakowski
چکیده

We report a case of a 73-year-old man, who was scheduled for elective dual chamber pacemaker implantation because of paroxysmal second-degree atrioventricular block. The pacemaker pocket was located in the left in-fraclavicular fossa. An active fixation ventricular lead was implanted into the right ventricular outflow tract (RVOT) through the left cephalic vein venesection, while another active fixation (Medtronic CapSure 5076MRI) atrial lead was located within the right atrial appendage using left subclavian vein puncture. During the procedure, due to anatomical inconveniences , several attempts at subclavian vein puncture were performed. Difficulties were also encountered during lead introduction through the superior vena cava. At the end of the procedure spontaneous dislocation of the atrial lead occurred. The young doctor repositioned this lead into the free wall of the right atrium (RA). Optimal sensing, pacing, and impedance parameters for both leads were obtained. The day after operation, the patient reported un-specific pain in the upper abdominal region and later strong, intermittent, stabbing pain in the lower chest. These symptoms intensified during inspiration and while lying on left side. The patient negated dyspnoea. On aus-cultation, normal vesicular breath sounds were heard. Neither pericardial fluid nor signs of RV perforation were found in transthoracic echocardiography (TTE). In the telemetric pacemaker assessment an increased pacing threshold of atrial lead was noticed (from < 0.5 V/0.4 ms during pacemaker implantation to 4.5 V/1 ms in unipolar pacing and 3 V/1 ms in bipolar pacing); no changes in sensing or impedance parameters were found. However, due to the lack of exact assessment of the atrial lead tip positioning, it was impossible to definitively exclude RA perforation (Figure 1 A). The device was reprogrammed to bipolar stimulation, leading to a decrease in reported complaints. On the next day a lack of atrial lead capture by threshold > 7.5 V/1.5 ms was recorded. Fluoroscopy revealed dislocation of the atrial lead, which was extracted subsequently. New lead was implanted through subclavian vein puncture into the right atrial appendage, obtaining good parameters (Figures 1 B, C). Several hours after reoperation the patient reported resting dyspnoea. Chest radiograph revealed large right-sided pneumothorax (Figure 1 D), which was cured by suction drainage. There are two probable causes of contralateral pneu-mothorax in the reported case. The first is pleura injury during Seldinger set introduction. Taking into account potential small mediastinal bleeding size, the venous system (especially the extrapericardial part of the superior vena cava) …

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Subacute pneumothorax contralateral to the venous access site associated with atrial lead perforation in a patient who was receiving corticosteroid therapy

Pneumothorax contralateral to the venous access site due to the right atrial lead is an uncommon complication. Concomitant steroid use is known as a risk factor of pacemaker lead perforation. We report a rare case of subacute contralateral pneumothorax due to a screw-in atrial lead perforation that occurred after dual-chamber pacemaker implantation in a patient who was receiving steroid therapy...

متن کامل

Erosive Pressure Dermatitis Associated with Cardiac Pacemaker

Erosive dermatitis is a skin reaction associated with implantable cardiac devices (ie, pacemakers and cardioverter defibrillators). We report a 65-year-old man with a history of several cerebrovascular attacks who developed an erythematous crusted erosive patch over the pacemaker site after implantation of two subsequent pacemakers. We suggest that physical irritation may be a trigger factor fo...

متن کامل

Leadless cardiac pacing

Transvenous cardiac pacemakers definitely improve quality of life and reduce mortality in at-risk patients, but they are associated with several potential device-related complications. Approximately 10% of patients experience complications related to transvenous implantation of the pacemaker. These may be attributable to either the pulse generator (hematoma, skin breakdown, pocket infection) or...

متن کامل

Complications of Pacemaker Implantation

Approximately 180000 patients undergo pacemaker implantation in the U.S each year [1]. In addition, the extreme elderly are the most rapidly growing segment of the U.S. [2,3] and pacemakers are commonly implanted in this population. There are reports of pacemaker implant complications (generally clinical trials reporting outcomes and incident complication rates) and fewer reports of complicatio...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2015