Failure of an implantable cardioverter-defibrillator to redetect ventricular fibrillation in patients with a nonthoracotomy lead system.

نویسندگان

  • W Jung
  • M Manz
  • R Moosdorf
  • B Lüderitz
چکیده

BACKGROUND Shock delivery of an implantable defibrillator may cause a change in the amplitude of endocardial electrograms and impair the detection of ventricular fibrillation. Thus, the effects of shock discharges on the amplitude of endocardial electrograms were evaluated in five patients undergoing implantation of a cardioverter-defibrillator in combination with a new nonthoracotomy lead system. METHODS AND RESULTS At implant, bipolar endocardial electrograms were recorded before each shock application, during ventricular fibrillation, during redetection of ventricular fibrillation in case the applied shock was ineffective, and at intervals of 5, 10, 20, 30, 60, and 120 seconds after each shock delivery. The amplitude of the endocardial electrograms decreased from 10.5 +/- 3.8 mV during sinus rhythm to 6.3 +/- 1.9 mV during initial ventricular fibrillation and declined to 2.2 +/- 1.3 mV during redetection of ventricular fibrillation. After successful termination, the following bipolar electrograms could be obtained at the predetermined intervals: 1.9 +/- 1.2 mV, 3.1 +/- 1.8 mV, 4.5 +/- 1.9 mV, 6.5 +/- 2.9 mV, 9.5 +/- 3.3 mV, and 10.4 +/- 3.8 mV. At predischarge testing, failure of redetection of ventricular fibrillation could be documented in two patients, requiring rescue external defibrillation in both cases to restore sinus rhythm. CONCLUSIONS These findings demonstrate that the implantable cardioverter-defibrillator did not ensure reliable redetection of ventricular fibrillation in patients using the implanted nonthoracotomy lead system. Thus, the potential risk of sudden cardiac death may persist in these patients despite defibrillator therapy.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Lead System

Background. Shock delivery of an implantable defibrillator may cause a change in the amplitude of endocardial electrograms and impair the detection of ventricular fibrillation. Thus, the effects of shock discharges on the amplitude of endocardial electrograms were evaluated in five patients undergoing implantation of a cardioverter-defibrillator in combination with a new nonthoracotomy lead sys...

متن کامل

Implantable cardioverter-defibrillator in a patient with dextrocardia situs inversus

Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...

متن کامل

A permanent transvenous lead system for an implantable pacemaker cardioverter-defibrillator. Nonthoracotomy approach to implantation.

A transvenous lead system for implantable defibrillators would obviate a surgical thoracotomy and reduce the morbidity and mortality associated with implantation. We evaluated the clinical performance of a new nonthoracotomy lead system that included a defibrillation lead in the coronary sinus. At the time of defibrillator implantation, transvenous defibrillation leads were inserted percutaneou...

متن کامل

Ensuring automatic detection of ventricular fibrillation.

T he sensitivity and specificity of ventricular fibrillation (VF) detection by automatic implantable cardioverter-defibrillators (ICDs) have received relatively little attention by clinical investigators compared with other aspects of defibrillator use. The paucity of reports on ventricular fibrillation detection is not related to its lack of importance but rather reflects how difficult it is t...

متن کامل

THE EFFECT OF NURSE-PEER-LED SUPPORT INTERVENTION ON QUALITY OF LIFE IN PATIENTS WITH AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR: A RANDOMIZED CLINICAL TRIAL

Background & Aims: Nowadays, the quality of life and how coping and treatment of patients with an implantable cardiac defibrillator has led to concern and attempts to reduce or eliminate the problems of these patients in the people who provide health care services. This study aimed to determine the effect of nurse-peer-led support intervention on quality of life in patients with an implantable ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 86 4  شماره 

صفحات  -

تاریخ انتشار 1992