A comparative in vitro hemolysis study of a pulsating intravenous artificial lung.

نویسندگان

  • Laura W Lund
  • Brack G Hattler
  • William J Federspiel
چکیده

A study was conducted to measure and compare the levels of hemolysis generated by an intravenous membrane oxygenation device referred to as the Intravenous Membrane Oxygenator (IMO) in previous literature. The device is comprised of several hundred hollow fiber membranes of approximately 40 cm in length that are woven in a fabric and wrapped around a centrally positioned balloon. The balloon, which is similar in shape and volume to an intra-aortic balloon, is rapidly inflated and deflated up to 300 bpm to augment gas exchange. To evaluate the hemolytic nature of this device, an in vitro test system was developed, consisting of two identical test loops, each incorporating a device test section of 1 inch in diameter, a heat exchanger, a Biomedicus pump head, a compliance bag, a venous reservoir bag, and Tygon tubing. Both loops were primed with 1.5 L of a bovine blood solution and run simultaneously at 37 degrees C for 6 hours at 4 L/min. Hematocrit and plasma free hemoglobin concentration were measured every 30 minutes to monitor hemolysis within each loop. This methodology was used to compare the hemolysis of the device at maximal pulsation with that of the control loop with an empty test section, as well as with a pulsing balloon of the same volume without any fibers. The results suggest that the hemolytic nature of the pulsating intravenous oxygenator is consistent with that of an intra-aortic balloon, a clinically used device not associated with any complications due to hemolysis.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An in vitro Comparative study upon the Hemolytic, Thrombogenic, Coagulation parameters and Stability properties of the Hemiscorpiuslepturus Venom

Hemiscorpius lepturus belonging to Hemiscorpiidae family is the most venomous of all types of scorpion existing in south west of Iran causing hemoglobinuria and dermal lesions by envenomation. We compare the hemolytic pattern upon time in different domestic animals upon time according to their different sphingomyelin contents. In addition other in vitro hematologic parameters, platelet lysis, c...

متن کامل

Preparation, properties and preclinical pharmacokinetics of low molecular weight heparin-modified isoliquiritigenin-loaded solid lipid nanoparticle

Low molecular weight heparin-modified isoliquiritigenin-loaded solid lipid nanoparticle (LMWH-ISL-SLN) was developed for injective application. The morphological observation, particle diameter and zeta potential of LMWH-ISL-SLN were characterized using transmission electron microscopy (TEM) and a Malvern Zetasizer. Its entrapment efficiency (EE) and drug loading (DL) were determined by ultracen...

متن کامل

Preparation, properties and preclinical pharmacokinetics of low molecular weight heparin-modified isoliquiritigenin-loaded solid lipid nanoparticle

Low molecular weight heparin-modified isoliquiritigenin-loaded solid lipid nanoparticle (LMWH-ISL-SLN) was developed for injective application. The morphological observation, particle diameter and zeta potential of LMWH-ISL-SLN were characterized using transmission electron microscopy (TEM) and a Malvern Zetasizer. Its entrapment efficiency (EE) and drug loading (DL) were determined by ultracen...

متن کامل

Preparation and in-vitro Evaluation of an Antisense-containing Cationic Liposome against Non-small Cell Lung Cancer: a Comparative Preparation Study

The current methods for treatment of cancers are inadequate and more specific methods such as gene therapy are in progress. Among different vehicles, cationic liposomes are frequently used for delivery of genetic material. This investigation aims to prepare and optimize DOTAP cationic liposomes containing an antisense oligonuclotide (AsODN) against protein kinase C alpha in non-small cells lung...

متن کامل

Preparation and in-vitro Evaluation of an Antisense-containing Cationic Liposome against Non-small Cell Lung Cancer: a Comparative Preparation Study

The current methods for treatment of cancers are inadequate and more specific methods such as gene therapy are in progress. Among different vehicles, cationic liposomes are frequently used for delivery of genetic material. This investigation aims to prepare and optimize DOTAP cationic liposomes containing an antisense oligonuclotide (AsODN) against protein kinase C alpha in non-small cells lung...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • ASAIO journal

دوره 48 6  شماره 

صفحات  -

تاریخ انتشار 2002