Absolute and Trend Accuracy of a New Regional Oximeter in Healthy Volunteers During Controlled Hypoxia

نویسندگان

  • Daniel Redford
  • Samata Paidy
  • Faisal Kashif
چکیده

BACKGROUND Traditional patient monitoring may not detect cerebral tissue hypoxia, and typical interventions may not improve tissue oxygenation. Therefore, monitoring cerebral tissue oxygen status with regional oximetry is being increasingly used by anesthesiologists and perfusionists during surgery. In this study, we evaluated absolute and trend accuracy of a new regional oximetry technology in healthy volunteers. METHODS A near-infrared spectroscopy sensor connected to a regional oximetry system (O3™, Masimo, Irvine, CA) was placed on the subject's forehead, to provide continuous measurement of regional oxygen saturation (rSO2). Reference blood samples were taken from the radial artery and internal jugular bulb vein, at baseline and after a series of increasingly hypoxic states induced by altering the inspired oxygen concentration while maintaining normocapnic arterial carbon dioxide pressure (PaCO2). Absolute and trend accuracy of the regional oximetry system was determined by comparing rSO2 against reference cerebral oxygen saturation (SavO2), that is calculated by combining arterial and venous saturations of oxygen in the blood samples. RESULTS Twenty-seven subjects were enrolled. Bias (test method mean error), standard deviation of error, standard error of the mean, and root mean square accuracy (ARMS) of rSO2 compared to SavO2 were 0.4%, 4.0%, 0.3%, and 4.0%, respectively. The limits of agreement were 8.4% (95% confidence interval, 7.6%-9.3%) to -7.6% (95% confidence interval, -8.4% to -6.7%). Trend accuracy analysis yielded a relative mean error of 0%, with a standard deviation of 2.1%, a standard error of 0.1%, and an ARMS of 2.1%. Multiple regression analysis showed that age and skin color did not affect the bias (all P > 0.1). CONCLUSIONS Masimo O3 regional oximetry provided absolute root-mean-squared error of 4% and relative root-mean-squared error of 2.1% in healthy volunteers undergoing controlled hypoxia.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Detection of critical cerebral desaturation thresholds by three regional oximeters during hypoxia: a pilot study in healthy volunteers

BACKGROUND Regional oximetry is increasingly used to monitor post-extraction oxygen status of the brain during surgical procedures where hemodynamic fluctuations are expected. Particularly in cardiac surgery, clinicians employ an interventional algorithm to restore baseline regional oxygen saturation (rSO2) when a patient reaches a critical desaturation threshold. Evidence suggests that monitor...

متن کامل

Efficacy of Tinospora cordifolia on Learning and Memory in Healthy Volunteers: A Double-Blind, Randomized, Placebo Controlled Study

Background. Tinospora cordifolia, an Indian medicinal plant, has been reported to have beneficial effects on disorders like peptic ulcer, hepatobiliary disorders, rheumatism, infectious diseases etc. It enhances cognition in normal rats and successfully overcomes cyclosporine induced memory deficit. Methods. Thirty healthy volunteers of age 18-30 years received Tinospora cordifolia (500 mg of p...

متن کامل

The accuracy and response characteristics of a simplified ear oximeter.

We evaluated the accuracy and speed of response of a newly available, lightweight, and relatively inexpensive ear oximeter (Biox II oximeter). The instrument was compared with another oximeter (Hewlett-Packard 47201A) under conditions of both steadily maintained and progressively increasing hypoxia induced in normal subjects by rebreathing. The new oximeter (Biox II), which can be operated in e...

متن کامل

The effects of motion on the performance of pulse oximeters in volunteers (revised publication).

BACKGROUND Pulse oximetry is considered a standard of care in both the operating room and the postanesthetic care unit, and it is widely used in all critical care settings. Pulse oximeters may fail to provide valid SpO2 data in various situations that produce low signal-to-noise ratio. Motion artifact is a common cause of oximeter failure and loss of accuracy. This study compares the accuracy a...

متن کامل

Effect of Nail Polish on Pulse Oximetry Findings in Healthy Volunteers: A Randomized Clinical Trail

Background and Objectives: Pulse oximetry is the most common technique for monitoring hemoglobin oxygen saturation (SpO2). Different colors and brands of nail polish may cause disturbance in the reading and interpretation of oxygen saturation. The aim of this study was to determine the effect of different colors of nail polish on oxygen saturation measured by pulse oximeter. Methods: Thirty he...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 119  شماره 

صفحات  -

تاریخ انتشار 2014