Phase calibration of hydrophones: heterodyne time-delay spectrometry and broadband pulse technique using an optical reference hydrophone

نویسندگان

  • Thomas L. Szabo
  • Ch. Koch
  • V. Wilkens
چکیده

Two hydrophone calibration techniques are presented which provide amplitude and phase of hydrophone sensitivity. Heterodyne time-delay spectrometry can be applied in the frequency range from 0.5 to 50 MHz to all common hydrophones but needs a standard to obtain absolute values. A pulse technique between 1 and 70 MHz allows exploiting of an optical multilayer hydrophone as a phase standard. In combination with HTDS, a calibration service is established that covers the complete spectrum of common piezoelectric and optical hydrophones for absolute amplitude and phase calibration. The measured complex frequency responses form the basis of deconvolution procedures suitable for the correction of measurements of, for example, the output of ultrasound diagnostic machines.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Heterodyne and Time-gated Time-delay Spectrometry for Amplitude and Phase Calibration of Hydrophones

The determination of amplitude and phase of the hydrophone sensitivity using two techniques based on time-delay spectrometry is presented. The frequency range in which the technique can be applied is from 1 MHz to 50 MHz. The agreement with the results obtained by alternative calibration techniques is excellent, which emphasises the correct working of this coherent technique. Introduction Calib...

متن کامل

A multiple-frequency hydrophone calibration technique.

A method is described for comparing the sensitivity of two hydrophones over the frequency range 1-15 MHz. This technique forms the basis for the dissemination of national ultrasonic standards in the U.K. over this frequency range. A reference hydrophone is placed in an ultrasonic field and then the device being calibrated is substituted and the two output voltages are compared. This substitutio...

متن کامل

Absolute Calibration of Hydrophones Using Heterodyne Interferometry and Zero-crossing Signal Demodulation

The current primary method for hydrophone calibration is the three-transducer spherical-wave reciprocity at NPL, covering the frequency range between 1 kHz up to 500 kHz with expanded uncertainties as low as ± 0.5 dB. Optical methods can provide an alternative method for primary calibration as they directly measure the acoustic particle velocity at a specific point in the field. The hydrophone ...

متن کامل

Two efficient methods for measuring hydrophone frequency response in the 100 kHz to 2 MHz range

As new medical applications of ultrasound emerge with operating frequencies in the hundreds of kilohertz to low megahertz region, it becomes more important to have convenient calibration methods for hydrophones in this frequency range. Furthermore, short diagnostic ultrasound pulses affected by finite amplitude distortion require that the hydrophone frequency response be known well below the ce...

متن کامل

The Calibration of Hydrophones by a Novel Optical Technique in the Frequency Range 10 Khz to 600 Khz

This paper presents a technique for the absolute calibration of hydrophones using a novel optical method. The method uses a thin, optically reflecting membrane suspended in the acoustic field to allow optical measurement of the acoustic particle velocity using a Doppler interferometer. The hydrophone under test is then substituted for the pellicle, and the hydrophone response to the known acous...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2004