Determination of Acoustic Power and Intensity Levels from Diagnostic Ultrasound Machines
نویسنده
چکیده
The average acoustic power and intensity levels from diagnostic ultrasound machines at five major Hospitals in Benue State were measured using ultrasound power meter UPM-DT-10AV. The ultrasound machine at Christian Hospital Mkar, Immaculate Conception Hospital Makurdi, Federal Medical centre Makurdi, Myom Hospital Gboko and Lobi Specialist Hospital Makurdi were used for the research work. The ultrasound power meter measures the out-put acoustic power from the ultrasound beams which were converted to intensity level, three set of readings were taken for each of the pre-timed duration of 5,10,15,20 and 25 minutes, the average values were then found. The results shows that the acoustic power and intensity levels from the measured ultrasound machines were not stable over time and that the intensity level increases exponentially with increase in the scanning duration. From the measured acoustic outputs, it was seen that the diagnostic ultrasound machine model EMP-860 located at Lobi Specialist Hospital Makurdi has the highest average intensity value of 0.51±0.01W/cm 2 for an exposed duration of 25 minutes and a minimum value of 0.06±0.003 W/cm 2 for an exposed duration of 5 minutes while the diagnostic ultrasound machine model SonoAce 8000 located at Federal Medical Centre Makurdi has the lowest measured average intensity value of 0.0067±0.0009 W/cm 2 and higher intensity value of 0.099±0.0012 W/cm 2 at an exposed duration of 5 and 25 minutes respectively. This work aims will assess ultrasound dosimetry compliance in our Hospitals and provides database for diagnostic ultrasound quality control in Benue State and Nigeria. The work has shown that some of the ultrasound machines have output intensities beyond the maximum safety limit of 0.094 W/cm 2 as recommended by WFUMB (1998) and FDA (1993) for fetal and abdominal scan.
منابع مشابه
Numerical Study for Optimizing Parameters of High-Intensity Focused Ultrasound-Induced Thermal Field during Liver Tumor Ablation: HIFU Simulator
Introduction High intensity focused ultrasound (HIFU) is considered a noninvasive and effective technique for tumor ablation. Frequency and acoustic power are the most effective parameters for temperature distribution and the extent of tissue damage. The aim of this study was to optimize the operating transducer parameters such as frequency and input power in order to acquire suitable temperatu...
متن کاملLow intensity ultrasound mechanical index as a parameter affecting of ability of proliferation and collagene type 1 expression of cells
Introduction: Mechanical index (MI) is used for quantifying acoustic cavitation and the relationship between acoustic pressure and the frequency. In this study, modeling of the MI was applied to provide treatment protocol and to understand the effective physical processes on reproducibility and gene expression of fibroblast cells. Skin damage can occur of burn, cuts, abrasions ...
متن کاملEffect of Ultrasonication on Membrane Structure and Flux Recovery for Whey Ultrafiltration
In this study, the effect of ultrasound irradiation on the flux recovery and fouling mitigation for the membranes made of the polysulfone by the phase inversion method were investigated. Two ultrasound irradiation regimes, including inside and outside the module, were chosen for this study. The experiments were conducted to investigate the effect of ultrasound irradiation on the membrane struct...
متن کاملBlood Brain Barrier Disruption by Focused Ultrasound and Microbubbles: A Numerical Study on Mechanical Effects
Introduction: Microbubbles are widely used as contrast agent in diagnostic ultrasound. Recently they have shown good potential for applications in the therapeutic field such as drug delivery to the brain. Recent studies have shown focused ultrasound in conjunction with injected micro-bubbles could temporarily disrupt blood-brain barrier and let therapeutic agents transport into...
متن کاملEvaluating the Effects of Dual Frequency Sonication Parameters on Acoustic Cavitation by Chemical Dosimeter Using Iodide
Background and Aims: Production of acoustic cavitation by sonication has been recently recommended as a targeted treatment. The experimental results from studies indicate that the activity of cavitation generated by bi- or multi-frequency ultrasound irradiation is higher than that caused by single frequency irradiation. In this study, effects dual (1 MHz and 40 kHz) and single frequency soni...
متن کامل