The open circuit nitrogen washout technique for measuring the lung volume in infants: methodological aspects.
نویسنده
چکیده
BACKGROUND Lung volume measurement by nitrogen washout is widely used in infants, though a lack of accuracy and changes of calibration over time have been reported. The potential sources of error were explored in order to increase the accuracy and reliability of the technique. METHODS A commercial system for nitrogen washout and a 0.5 litre calibrating syringe as a lung model were used to perform over 2000 in vitro washouts, including simulated rapid breathing, shallow breathing, periodic breathing, sighs, and brief apnoeas. A constant 10 l/min bias flow of oxygen and extended equipment warming times were employed. A collapsible breathing bag was incorporated into the washout circuit. Following a single two point calibration, known air volumes from 42 ml to 492 ml were measured by nitrogen washout over a 14 hour period. The flow waveform in the nitrogen mixing chamber during a washout in vitro, with and without the breathing bag in the circuit, was also studied. RESULTS The mean coefficient of variation of all volumes was 0.66%. The mean difference between measured and known volumes was 0.30 ml (95% confidence interval (CI) -0.18 to 0.79). This difference was not statistically significant (p = 0.22). The mean percentage error was -0.1% (range -0.47% to 0.46%). Nitrogen calibration remained stable for 14 hours. Without the breathing bag flow transients were frequent in the mixing chamber during in vitro washout. CONCLUSIONS This technique increases the accuracy in vitro and the precision in vivo of volume measurement by nitrogen washout. Sources of potential errors including baseline drifting and inadequate equipment warming times were identified. The breathing bag acted as a buffer reservoir, preventing large swings in flows within the nitrogen mixing chamber during washouts, and should be an integral component of the nitrogen washout circuit.
منابع مشابه
A novel non-invasive technique for measuring the residual lung volume by nitrogen washout with rapid thoracoabdominal compression in infants.
BACKGROUND The functional residual capacity (FRC), the only lung volume to be routinely measured in infants, is an unreliable volume landmark. In addition to FRC, the residual volume (RV) was measured by nitrogen washout using rapid thoracoabdominal compression (RTC) in nine infants with cystic fibrosis aged 5-31 months. METHODS A commercial system for nitrogen washout to measure lung volumes...
متن کاملSeries "standards for Infant Respiratory Function Testing: Ers/ats
The bias flow nitrogen washout technique for measuring the functional residual capacity in infants. M.G. Morris, P. Gustafsson, R. Tepper, M. Gappa, J. Stocks, on behalf of the ERS/ATS Task Force on Standards for Infant Respiratory Function Testing. #ERS Journals Ltd 2001. ABSTRACT: The functional residual capacity (FRC) is the most commonly measured static lung volume in infants. It is importa...
متن کاملMoment ratio analysis of multiple breath nitrogen washout in infants with lung disease.
Measurement of lung volumes at end expiratory level and assessment of ventilation inhomogeneity is important for respiratory management in infants with lung disease. This study compared multiple breath nitrogen washout was compared with body plethysmography to measure functional residual capacity in infants and assessed ventilation inhomogeneity using mean dilution numbers and alveolar based ga...
متن کاملDefining the appropriate waiting time between multiple-breath nitrogen washout measurements.
Static lung volume measurements allow objective assessment of total lung capacity (TLC), functional residual capacity (FRC) and residual volume, and assist in the diagnosis and management of lung disease [1, 2]. The use of gas dilution techniques, such as multiple-breath nitrogen washout (MBNW), require minimal patient cooperation and allow FRC to be determined in those individuals unable to co...
متن کاملNitrogen washout/washin, helium dilution, and computed tomography in the assessment of end-expiratory lung volume
INTRODUCTION End expiratory lung volume (EELV) measurement in the clinical setting is routinely performed using the helium dilution technique. A ventilator that implements a simplified version of the nitrogen washout/washin technique is now available. We compared the EELV measured by spiral computed tomography (CT) taken as gold standard with the lung volume measured with the modified nitrogen ...