Radiation survey around a Liac mobile electron linear accelerator for intraoperative radiation therapy

نویسندگان

  • Mario Ciocca
  • Guido Pedroli
  • Roberto Orecchia
  • Andrea Guido
  • Federica Cattani
  • Raffaella Cambria
  • Umberto Veronesi
چکیده

The aim of this study was to perform a detailed analysis of the air kerma values around a Liac mobile linear accelerator working in a conventional operating room (OR) for IORT. The Liac delivers electron beams at 4, 6, 8 and 10 MeV. A radiation survey to determine photon leakage and scatter consisted of air kerma measurements on a spherical surface of 1.5 m radius, centered on the titanium exit window of the accelerating structure. Measurements were taken using a 30 cm3 calibrated cylindrical ion chamber in three orthogonal planes, at the maximum electron energy. For each point, 10 Gy was delivered. At selected points, the quality of x-ray radiation was determined by using lead sheets, and measurements were performed for all energies to investigate the energy dependence of stray radiation. The photon scatter contribution from the metallic internal patient-shielding in IORT, used to protect normal tissues underlying the target, was also evaluated. At seven locations outside the OR, the air kerma values derived from in-room measurements were compared to measurements directly performed using a survey meter. The results, for a delivered dose of 10 Gy, showed that the air kerma values ranged from approximately 6 microGy (upper and rear sides of the Liac) to 320 microGy (lateral to beam stopper) in the two orthogonal vertical planes, while values lower than 18 microGy were found in the horizontal plane. At 10 MeV, transmission behind 1 cm lead shield was found to be 42%. The use of internal shielding appeared to increase the photon scatter only slightly. Air kerma values outside the OR were generally lower than 1 mGy for an annual workload of 200 patients. Thus, the Liac can safely work in a conventional OR, while the need for additional shielding mainly depends on patient workload. Our data can be useful for centers planning to implement an IORT program using a mobile linear accelerator, permitting radiation safety personnel to estimate in advance the shielding required for a particular workload.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator

Background: One of the main problems of dedicated IORT accelerators is to determine dosimetric characteristics of the electron beams. Monte Carlo simulation of IORT accelerator head and produced beam will be useful to improve the accuracy of beam dosimetry.Materials and Methods: Liac accelerator head was modeled using the BEAMnrcMonte Carlo simulation system. Phase-space files were generated at...

متن کامل

اعتبارسنجی سر شتاب‌دهندۀ پرتودرمانی حین عمل‌ جراحی LIAC با استفاده از روش مونت‌کارلو

With the development of science and technology in the medical field and need more than ever to accuracy and quality of treatment has been causer increase different radiotherapy methods. Almost the past two decade’s intraoperative radiation therapy (IORT) as one of the new techniques used to treat cancer patients. One of the problems with this method is obtaining accurate dosimetry, becaus...

متن کامل

Commissioning the First Mobile Dedicated Accelerator for Intraoperative Electron Radiotherapy in Iran

Introduction: Intraoperative radiotherapy is a radiotherapy technique in which a high single fraction of radiation dose is delivered to the patient after surgery and Concurrent with anesthesia time. The most frequent method for IORT implementation is Intraoperative electron radiotherapy (IOERT), in which, some dedicated and high dose per pulse electron accelerators are employe...

متن کامل

Dosimetric characteristics of electron beams produced by two mobile accelerators, Novac7 and Liac, for intraoperative radiation therapy through Monte Carlo simulation

The Novac7 and Liac are linear accelerators (linacs) dedicated to intraoperative radiation therapy (IORT), which produce high energy, very high dose-per-pulse electron beams. The characteristics of the accelerators heads of the Novac7 and Liac are different compared to conventional electron accelerators. The aim of this work was to investigate the specific characteristics of the Novac7 and Liac...

متن کامل

Physical characteristics of electron beam from conventional and beam shaper IOERT applicator: A comparison study

Introduction: Intraoperative electron radiation therapy (IOERT) is one of the cancer treatment techniques that delivers high doses to tumor bed during surgery. IOERT can be performed by either conventional LINACs or dedicated IORT accelerators such as LIAC (Light Intraoperative Accelerator). Two types of applicators can be used with LIAC dedicated accelerator including conventi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2009