Neural response thresholds in the Nucleus Contour cochlear implant before and after stylet removal.

نویسندگان

  • Robinson Koji Tsuji
  • Maria Valéria Schmidt Goffi-Gomez
  • Cristina Ornelas Peralta
  • Mariana Cardoso Guedes
  • Ana Tereza Magalhães
  • Rubens Brito Neto
  • Ricardo Ferreira Bento
چکیده

CONCLUSION The study shows that there are differences in the measurement of the action potentials with and without the stylet in the Nucleus Freedom Contour Advance that are higher in the apex than in the base of the cochlea. OBJECTIVES To determine if there are differences in the intraoperative impedances and in the neural response telemetry threshold values in the Nucleus Freedom Contour Advance before and after stylet removal. SUBJECTS AND METHODS This was a prospective clinical study. Intraoperative impedances and neural response telemetry in users of the Freedom Contour Advance Cochlear Implant were measured before and after stylet removal. RESULTS There was a significant reduction in the impedance values of an average 1.5 kOmega+/-2.3 in common ground mode and 1.3 kOmega+/-2.3 for all monopolar modes after the stylet removal (p < 0.001). When analyzing the apical, medium, and basal electrodes, there was a statistically significant reduction in the neural response thresholds after stylet removal only in the apical electrodes (p = 0.001).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Relationship between Electrically Evoked Compound Action Potential Thresholds and Auditory, Language, and Speech Progress after Cochlear Implant Surgery

Introduction: Electrically evoked compound action potential (ECAP) is an objective auditory response that can be used in the programing of cochlear implants. The aims of this study were to monitor ECAP thresholds and auditory, language and speech progress for 6 months after cochlear implant surgery and to evaluate any relationship between them.   Materials and Methods: Ten children with a m...

متن کامل

Evaluation of Saccular Function Pre-Post Cochlear Implant Surgery Using VEMPs

Objectives: The aim of present study was evaluation of saccule function in cochlear implant candidates with severe to profound sensory neural hearing loss Before and after cochlear implantation Methods: In this study 35 cochlear implant (CI) candidates with bilateral severe to profound sensory neural hearing loss before and about 30 days after cochlear implant and 20 normal-hearing cases as ...

متن کامل

Perimodiolar electrode position: Effects on thresholds, comfort levels, impedance measurements, and neural response telemetry

OBJECTIVE: This multicenter study describes the effects of perimodiolar electrode position on 50 adult patients (31 with an implanted Nucleus 24 Contour perimodiolar hugging electrode array [Cochlear Limited, Lane Cove, NSW, Australia] and 19 with an implanted straight electrode array) with a severe-to-profound sensorineural hearing loss. MATERIALS AND METHODS: In all patients, intraoperative i...

متن کامل

Evaluation of the Slope of Amplitude Growth Function Changes of the Electrically Evoked Action Potential in Three Months after Receiving the Device in Children with Cochlear Implant

Objective: In neural response telemetry, intracochlear electrodes stimulate the auditory nerve and record the neural responses. The electrical stimulation is sent to the auditory nerve by an electrode and the resulted response, called electrically evoked compound action potential, is recorded by an adjacent electrode. The most important clinical applications of this test are evaluation and moni...

متن کامل

A Psychophysics experimental software to evaluate electrical pitch discrimination in Nucleus cochlear implanted patients

Multichannel electrode array design in cochlear implants has evolved into two major categories: straight and perimodiolar electrodes. When implanted, the former lies along the outer wall of the scala tympani, while the later are located closer to the modiolus, where the neural ends are. Therefore, a perimodiolar position of the electrode array could be expected to result in reduced stimulus thr...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Acta oto-laryngologica

دوره 129 11  شماره 

صفحات  -