SU-E-T-648: Comparison of VMAT Vs Arc Treatment Plans for Patients Undergoing SBRT of Early-Stage Non-Small Cell Lung Cancer (NSCLC).

نویسندگان

  • T Neicu
  • P Chan
  • V Valakh
  • M Bizhan
  • C Miyamoto
  • S Li
چکیده

PURPOSE To evaluate the dosimetric implications of using VMAT (Volume Modulated Arc Therapy) treatment planning techniques compared to traditional Arc therapy methods for patients undergoing SBRT (Stereotactic Body Radiation Therapy) for early-stage Non-Small Cell Lung Cancer (NSCLC). METHODS Ten NSCLC cancer patients are planned with both VMAT and Arc techniques. The SBRT treatment plans comparison was quantified by several Dose-Volume Histogram (DVH) indicators including mean, maximum and minimum doses for GTV, ITV, PTV, OAR (Organs At Risk), and V95 (volume receiving at least 95% of the prescribed dose) for PTV. RESULTS On average VMAT plans require for treatment delivery 16.6 ± 20.2 % more monitor units (MU) than the traditional Arc plans. The average PTV minimum, maximum and mean doses as a percentage of prescribed dose are 94.5 ± 3.9 %, 114.1 ± 3.3 % and 106.6 ± 1.6 % for VMAT vs 91.6 ± 4.4 %, 119.5 ± 5.3 % and 109.5 ± 2.5 % for the Arc technique. The V95 PTV coverage for VMAT plans range from 99.4 % to 100 % with a mean of 99.7 %, compared with a range of 96.8 % to 100 % with a mean of 99 % for the Arc plans. The maximum dose received by the lungs, spinal cord and chest wall show on average significant increases for Arc plans as opposed to VMAT plans (5.7 ± 6 % increase for lungs, 4.4 ± 9.2 % for cord and 2.4 ± 6.3 % for chest wall). The average mean doses and minimum doses for the OAR are similar for both techniques. CONCLUSIONS The comparison of VMAT vs Arc plans for SBRT of NSCLC patients is subject to many variables, including GTV and PTV volume sizes, shape and their proximity relative to the OAR.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

SU-E-T-254: Quantitative Evaluation of Three Mathematical Models for Radiation-Induced Secondary Cancer Risk for Lung in Stereotactic Body Radiotherapy.

PURPOSE Several mathematical models exist to predict secondary cancer risks from radiotherapy treatment. In this study, we aimed to compare three dosimetry-based models in the application of stereotactic body radiotherapy (SBRT) of early-stage non-small cell lung cancer (NSCLC) with three radiotherapy treatment modalities. METHODS Ten patients who received definitive SBRT for early-stage NSCL...

متن کامل

A dosimetric comparison of three‐dimensional conformal radiotherapy, volumetric‐modulated arc therapy, and dynamic conformal arc therapy in the treatment of non‐small cell lung cancer using stereotactic body radiotherapy

This study evaluates three-dimensional conformal radiotherapy (3D CRT), volumetric-modulated arc therapy (VMAT), and dynamic conformal arc therapy (DCAT) planning techniques using dosimetric indices from Radiation Therapy Oncology Group (RTOG) protocols 0236, 0813, and 0915 for the treatment of early-stage non-small cell lung cancer (NSCLC) using stereotactic body radiotherapy (SBRT). Twenty-fi...

متن کامل

Comprehensive dosimetric planning comparison for early‐stage, non‐small cell lung cancer with SABR: fixed‐beam IMRT versus VMAT versus TomoTherapy

Volumetric-modulated arc therapy (VMAT) is emerging as a leading technology in treating early-stage, non-small cell lung cancer (NSCLC) with stereotactic ablative radiotherapy (SABR). However, two other modalities capable of deliver-ing intensity-modulated radiation therapy (IMRT) include fixed-beam and helical TomoTherapy (HT). This study aims to provide an extensive dosimetric compari-son amo...

متن کامل

A dosimetric evaluation of VMAT for the treatment of non‐small cell lung cancer

The purpose of this study was to demonstrate the dosimetric potential of volumetric-modulated arc therapy (VMAT) for the treatment of patients with medically inoperable stage I/II non-small cell lung cancer (NSCLC) with stereotactic body radiation therapy (SBRT). Fourteen patients treated with 3D CRT with varying tumor locations, tumor sizes, and dose fractionation schemes were chosen for study...

متن کامل

A review of stereotactic body radiotherapy – is volumetric modulated arc therapy the answer?

Stereotactic body radiotherapy (SBRT) is a high precision radiotherapy technique used for the treatment of small to moderate extra-cranial tumours. Early studies utilising SBRT have shown favourable outcomes. However, major disadvantages of static field SBRT include long treatment times and toxicity complications. Volumetric modulated arc therapy (VMAT) and intensity modulated radiotherapy (IMR...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 39 6Part20  شماره 

صفحات  -

تاریخ انتشار 2012