False-positive cryptococcal antigen test associated with use of BBL Port-a-Cul transport vials.
نویسندگان
چکیده
A total of 52 residual CSF and serum specimens, which were originally negative with the Cryptococcal Antigen Latex Agglutination System (CALAS), were shown to become falsely positive after placement in BBL Port-A-Cul anaerobic transport vials. This transport device, although excellent for specimen transportation for subsequent culture, should not be used if cryptococcal antigen testing is needed.
منابع مشابه
Comparison of three transport systems (Starplex StarSwab II, the new Copan Vi-Pak Amies Agar Gel collection and transport swabs, and BBL Port-A-Cul) for maintenance of anaerobic and fastidious aerobic organisms.
Of utmost importance in evaluations of clinical samples for infectious agents is proper specimen transport to the clinical laboratory. In the present study we compared three transport systems (the new Starplex StarSwab II, the new Copan Vi-Pak Amies Agar Gel collection and transport swabs, and the BBL Port-A-Cul) for survival of anaerobic and fastidious aerobic bacteria. The new Copan Vi-Pak sy...
متن کاملEvaluation of Port-A-Cul transport system for protection of anaerobic bacteria.
The protection of anaerobes in Port-A-Cul (PAC) transport system (Bioquest, Div. of Becton, Dickinson &Co., Cockeysville, Md.) tubes and vials was studied. Ten species of obligately anaerobic bacteria commonly isolated from clinical specimens were used to prepare simulated swab and fluid specimens in high and low concentrations. Samples in PAC tubes and vials were held for 2, 24, and 48 h at am...
متن کاملNew cause for false-positive results with the cryptococcal antigen test by latex agglutination.
The highly specific and sensitive latex agglutination test for cryptococcal antigen detection in cerebrospinal fluid is routine in many hospitals. Contamination of cerebrospinal fluid by a minute amount of syneresis fluid (surface condensation) from agar gave a strongly positive reaction which was heat stable, was not eliminated by pronase treatment, and was not detected by the normal rabbit gl...
متن کاملFalse-negative cerebrospinal fluid cryptococcal latex agglutination tests for patients with culture-positive cryptococcal meningitis.
Three cases of false-negative cerebrospinal fluid latex agglutination test results for patients with culture-positive cryptococcal meningitis are reported. False-negative results occurred in settings of low cryptococcal antigen concentrations in cerebrospinal fluid and were dependent on the latex agglutination test kit used. Investigation of each case revealed that prozone phenomena or interfer...
متن کاملInactivated pronase as the cause of false-positive results of serum cryptococcal antigen tests.
Four patients who had acquired immunodeficiency syndrome and who were evaluated for headache within a 3-week period had false-positive results of serum cryptococcal antigen tests. This cluster of false-positive test results appeared to be due to inactivation of the pronase vial in the test kit, a cause that has not been reported previously.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical microbiology
دوره 49 2 شماره
صفحات -
تاریخ انتشار 2011