Initial experience in Brazil with the Helex septal occluder for percutaneous occlusion of atrial septal defects.
نویسندگان
چکیده
OBJECTIVE To evaluate the initial clinical experience with the Helex septal occluder for percutaneous closure of atrial septal defects. METHODS Ten patients underwent the procedure, 7 patients with ostium secundum atrial septal defects (ASD) with hemodynamic repercussions and 3 patients with pervious foramen ovale (PFO) and a history of stroke. Mean age was 33.8 years and mean weight was 55.4 kg. Mean diameter by transesophageal echocardiography and mean stretched ASD diameter were 11.33 3.3mm, and 15.2 3.8mm, respectively. The Qp/Qs ratio was 1.9 0.3 in patients with ASD. RESULTS Eleven occluders were placed because a patient with 2 holes needed 2 devices. It was necessary to retrieve and replace 4 devices in 3 patients. We observed immediate residual shunt (< 2mm) in 4 patients with ASD, and in those with patent foramen ovale total occlusion of the defect occurred. No complications were noted, and all patients were discharged on the following day. After 1 month, 2 patients with ASD experienced trivial residual shunts (1mm). In 1 patient, we observed mild prolapse in the proximal disk in the right atrium, without consequences. CONCLUSION The Helex septal occluder was safe and effective for occluding small to moderate atrial septal defects. Because the implantation technique is demanding, it requires specific training of the operator. Even so, small technical failures may occur in the beginning of the learning curve, but they do not involve patient safety.
منابع مشابه
Urgent Surgical Intervention for Embolized Cardiac Occluder Devices: A Case Series
Introduction: In this study, we sought to illustrate our experience in urgent surgical management for embolized cardiac septal occlude devices resulting from trans-catheter closure of atrial septal defect and ventricular septal defect. Mathrials and Methods: We retrospectively reviewed four patients aged 2–10 years who underwent urgent surgery due to cardiac septal occluder embolization between...
متن کاملResults of the U.S. multicenter pivotal study of the HELEX septal occluder for percutaneous closure of secundum atrial septal defects.
OBJECTIVES This study sought to compare the safety and efficacy of the HELEX septal occluder (HSO) with surgical repair of atrial septal defect (ASD). BACKGROUND The HSO is a low-profile, double-disk occluder device for percutaneous closure of secundum ASD. METHODS Patients were enrolled (HSO arm prospectively, surgery arm prospectively/retrospectively) from 14 U.S. sites and followed up fo...
متن کاملDevelopment and testing of the Helex septal occluder, a new expanded polytetrafluoroethylene atrial septal defect occlusion system.
BACKGROUND A variety of transcatheter atrial septal defect (ASD) occluders are currently in use around the world. Although for the most part effective, all of these devices lack features that would be desirable in a "perfect" device. The Helex septal occluder is a new type of device designed to improve the results of transcatheter ASD closure. This study was designed to examine the effectivenes...
متن کاملRecent advances in managing septal defects: atrial septal defects
The purpose of this review is to discuss the management of atrial septal defects (ASD), paying particular attention to the most recent developments. There are four types of ASDs: ostium secundum, ostium primum, sinus venosus, and coronary sinus defects. The fifth type, patent foramen ovale-which is present in 25 to 30% of normal individuals and considered a normal variant, although it may be th...
متن کامل[Percutaneous closure of large atrial septal defects].
BACKGROUND Percutaneous occlusion of secundum type atrial septal defect with an Amplatzer device is an effective treatment alternative. However, there is little information about the use of this treatment for large defects. PATIENTS AND METHOD We retrospectively analyzed the initial and follow-up findings in 31 patients aged 41 18 years who had large atrial septal defects (>= 30 mm diameter)....
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Arquivos brasileiros de cardiologia
دوره 81 5 شماره