Outlet strut fracture of Björk-Shiley convexo concave heart valves: the UK cohort study.

نویسندگان

  • R Z Omar
  • L S Morton
  • D A Halliday
  • E M Danns
  • M T Beirne
  • W J Blot
  • K M Taylor
چکیده

OBJECTIVE To investigate the risk of outlet strut fracture (OSF) in Björk-Shiley convexo concave (BSCC) valves in relation to patients' clinical and valve characteristics. DESIGN A cohort of 2977 patients with 3325 valves with a follow up of 18 years. SETTING 38 cardiac implantation centres in the UK. RESULTS 56 OSF events were reported with 43 occurring in mitral and 13 in aortic valves. The overall OSF rate was 0.17%/year. No dominant clinical factor of risk was found, but multiple regression analysis identified age, body surface area, valve size, shop order fracture rate, and manufacturing period as risk factors for OSF. A 4% (95% confidence interval (CI) 2% to 6%) decrease in the risk of OSF was observed for each advancing year of age and a fivefold (95% CI 2 to 13) increase in risk for a 0.5 m(2) increase in body surface area. The association between the risk of OSF and valve size was not constant over time. Excess risks among 31 mm and 33 mm sizes (mainly mitral valves) decreased over time while that for 23 mm (almost all aortic valves) increased. The risk of OSF increased by 40% (95% CI 20% to 50%) for a unit increase in the fracture rate of other valves in the same batch. For valves manufactured during 1981 to 1984 the risk of OSF was 4 (95% CI 2 to 12) times greater than for valves manufactured before 1981. CONCLUSIONS The OSF rates for 60 degrees BSCC valves observed in the UK are the highest among all monitored populations. The changing patterns of mitral and aortic valve OSF rates over time observed in this study have not been identified previously and highlight the need for continued monitoring of patients with the BSCC valve.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Decision guidelines for prophylactic replacement of Björk-Shiley convexo-concave heart valves: impact on clinical practice.

BACKGROUND Because of risk of outlet strut fracture, prophylactic replacement should be considered for Björk-Shiley convexo-concave (BScc) valve recipients. We assessed the effects of epidemiological and decision-analytic guidelines on actual BScc valve replacement. METHODS AND RESULTS We performed a retrospective cohort study including all 2263 Dutch BScc patients with a mean follow-up of 11...

متن کامل

The Björk-Shiley convexo-concave heart valve experience from the perspective of the supervisory panel.

The 20-year activities of a medical supervisory panel appointed under the terms of a settlement agreement of the Bowling v. Pfizer class action suit involving the Björk-Shiley convexo-concave (BSCC) heart valve are detailed. Of approximately 86,000 valves implanted, catastrophic failure of the valve was reported in 663 patients from 1978 to 2012. In 1994, a 7-member medical panel consisting of ...

متن کامل

Age thresholds for prophylactic replacement of Björk-Shiley convexo-concave heart valves. A clinical and economic evaluation.

BACKGROUND Björk-Shiley convexo-concave heart valves have an increased risk of mechanical failure. One might consider prophylactic rereplacement as a preventive measure to avert the disastrous consequences of these failures. We investigated the effect that prophylactic rereplacement has on survival of individual patients and on the medical costs. METHODS AND RESULTS Quantitative estimates for...

متن کامل

Psychological distress among recipients of Björk-Shiley convexo-concave valves: the impact of information.

OBJECTIVE To establish whether there is more psychological distress among recipients of Björk-Shiley convexo-concave (BScc) valves than among recipients of other valves not known to fracture. DESIGN Cross sectional study. PATIENTS 137 patients who received either a BScc valve or a Sorin Biomedical spherical valve during a randomised trial between 1982 and 1983 at St Antonius Hospital, Nieuw...

متن کامل

Prophylactic replacement of Bjŏrk-Shiley convexo-concave heart valves: an easy-to-use tool to aid decision-making in individual patients.

OBJECTIVE To develop an easy-to-use tool for decision-making on prophylactic replacement of Björk-Shiley convexo-concave heart valves. DESIGN Decision analysis comparing elective replacement with observation. SETTING Referral centres for patients with artificial heart valves. PATIENTS Quantitative estimates were obtained from a follow up study conducted in The Netherlands, including 2303 ...