Assessment of the diagnostic accuracy of the TDx-FLM II to predict fetal lung maturity.
نویسندگان
چکیده
BACKGROUND Because respiratory distress syndrome (RDS) affects 1% of live births, accurate and rapid assessment of markers of fetal lung maturity is critical to clinicians in deciding whether to deliver a preterm infant. Our objective was to determine the optimal diagnostic cutoff value for the TDx-FLM II assay (Abbott Laboratories) for predicting clinically significant RDS. METHODS Amniotic fluid TDx-FLM II data were collected retrospectively over 4 years. Women were included in the study if they had delivered within 72 h of TDx-FLM II testing and both the mother and infant charts could be reviewed. Women who had been treated with steroids and delivered unaffected infants were excluded from the analysis. The diagnosis of RDS was defined as infants who either were treated with surfactant and/or were placed on a ventilator and/or required continuous positive airway pressure for >1 day. RESULTS A total of 185 women met all entry criteria (15 RDS, 170 non-RDS). A cutoff value for a mature result of >or=45 mg/g gave a sensitivity of 100% (95% confidence interval, 82-100%) and a specificity of 90% (95% confidence interval, 78-89%). CONCLUSIONS The TDx-FLM II appears to predict clinically significant RDS when a cutoff of >or=45 mg/g is used for mature results. Further studies will be required to confirm these findings.
منابع مشابه
Validation of lamellar body counts using three hematology analyzers.
The lamellar body count (LBC) represents an alternative method to the TDx-FLM II (Abbott Laboratories, Abbott Park, IL), which is planned to be discontinued, for assessing fetal lung maturity. Our objective was to validate the LBC on 3 hematology analyzers (Coulter LH 750 and Coulter Ac.T diff2, Beckman Coulter, Brea, CA; and Sysmex XE-2100, Sysmex, Mundelein, IL) to serve as a template for oth...
متن کاملPreanalytical factors that influence the Abbott TDx Fetal Lung Maturity II assay.
BACKGROUND The TDx Fetal Lung Maturity II (FLM II) assay uses amniotic fluid to assess lung maturity of the unborn infant. We investigated common preanalytical factors that influence FLM results, including centrifugation, sample storage, and contamination by whole blood. METHODS We tested 18 specimens after centrifugation and after resuspension by vortex-mixing. We also analyzed 23 specimens ...
متن کاملAnalytical and clinical evaluation of refractive index-matched anomalous diffraction (RIMAD) for assessment of fetal lung maturation.
We have evaluated refractive index-matched anomalous defraction (RIMAD) (Dubin SB, Clin Chem 1988;34:938-43) as a potential method for assessment of fetal lung maturity (FLM). This method determines the total light scattered by the surfactant-containing lamellar bodies by subtraction of the A650 from amniotic fluid diluted in glycerol from that of amniotic fluid diluted in distilled water. It i...
متن کاملEvaluation of the Amniostat-FLM assay for assessment of fetal lung maturity.
Results of the "Amniostat-FLM" assay, a rapid semiquantitative test for phosphatidylglycerol, were compared with determinations of the lecithin/sphingomyelin (L/S) ratio, with phosphatidylglycerol measured by two-dimensional thin-layer chromatography, and with results of the "shake test" for 94 specimens of amniotic fluid. Correlation between results with the Amniostat and the other tests was e...
متن کاملDecreased laboratory testing for lecithin-to-sphingomyelin ratio and phosphatidylglycerol after fetal lung maturity assessment from lamellar body count in amniotic fluid.
The objectives of this study were to do inexpensive lamellar body count (LBC) in amniotic fluid, to do statistical analysis to evaluate cutoff values for fetal lung maturity (FLM) and fetal lung immaturity (FLI), to derive a threshold for obtaining a lecithin-to-sphingomyelin (L/S) ratio and phosphatidylglycerol percentage (%PG), and to determine the potential cost savings to the hospital if th...
متن کاملذخیره در منابع من