In Heart Failure Patients with Left Bundle Branch Block Single Lead MultiSpot Left Ventricular Pacing Does Not Improve Acute Hemodynamic Response To Conventional Biventricular Pacing. A Multicenter Prospective, Interventional, Non-Randomized Study
نویسندگان
چکیده
INTRODUCTION Recent efforts to increase CRT response by multiSPOT pacing (MSP) from multiple bipols on the same left ventricular lead are still inconclusive. AIM The Left Ventricular (LV) MultiSPOTpacing for CRT (iSPOT) study compared the acute hemodynamic response of MSP pacing by using 3 electrodes on a quadripolar lead compared with conventional biventricular pacing (BiV). METHODS Patients with left bundle branch block (LBBB) underwent an acute hemodynamic study to determine the %change in LV+dP/dtmax from baseline atrial pacing compared to the following configurations: BiV pacing with the LV lead in a one of lateral veins, while pacing from the distal, mid, or proximal electrode and all 3 electrodes together (i.e. MSP). All measurements were repeated 4 times at 5 different atrioventricular delays. We also measured QRS-width and individual Q-LV durations. RESULTS Protocol was completed in 24 patients, all with LBBB (QRS width 171±20 ms) and 58% ischemic aetiology. The percentage change in LV+dP/dtmax for MSP pacing was 31.0±3.3% (Mean±SE), which was not significantly superior to any BiV pacing configuration: 28.9±3.2% (LV-distal), 28.3±2.7% (LV-mid), and 29.5±3.0% (LV-prox), respectively. Correlation between LV+dP/dtmax and either QRS-width or Q-LV ratio was poor. CONCLUSIONS In patients with LBBB MultiSPOT LV pacing demonstrated comparable improvement in contractility to best conventional BiV pacing. Optimization of atrioventricular delay is important for the best performance for both BiV and MultiSPOT pacing configurations. TRIAL REGISTRATION ClinicalTrials.gov NTC01883141.
منابع مشابه
His Bundle Pacing Or Biventricular Pacing For Cardiac Resynchronization Therapy In Heart Failure: Discovering New Methods For An Old Problem.
Heart failure (HF) is one of the biggest epidemics of modern cardiovascular medicine. Cardiac resynchronization therapy (CRT) with biventricular (BiV) pacing has proven to have an integral role in the management of patients with reduced left ventricular (LV) function and left bundle branch blocks (LBBB). However, CRT with BiV pacing is not always feasible and even when it is, the percentage of ...
متن کاملElectrical and mechanical components of dyssynchrony in heart failure patients with normal QRS duration and left bundle-branch block: impact of left and biventricular pacing.
BACKGROUND Resynchronization pacing is an effective symptomatic treatment for heart failure patients with prolongation of the QRS duration (QRSd). Dyssynchronous contraction of the left ventricle is also observed with normal QRSd. We set out to determine how electrical activation of the left ventricular (LV) free wall differed between patients with left bundle-branch block (LBBB) and normal QRS...
متن کاملLeft ventricular endocardial or triventricular pacing to optimize cardiac resynchronization therapy in a chronic canine model of ischemic heart failure.
Cardiac resynchronization therapy (CRT) is a proven treatment for heart failure but ~30% of patients appear to not benefit from the therapy. Left ventricular (LV) endocardial and multisite epicardial [triventricular (TriV)] pacing have been proposed as alternatives to traditional LV transvenous epicardial pacing, but no study has directly compared the hemodynamic effects of these approaches. Le...
متن کاملThe successful treatment of end stage of heart failure associated with complete right bundle branch block with biventricular pacemaker placement.
Biventricular pacing has been used to treat patients with symptomatic heart failure, systolic left ventricular dysfunction and intraventricular conduction delays. This modality is usually reserved for the treatment of patients with a left bundle branch block pattern on electrocardiogram. We report the successful use of biventricular pacing in a patient with heart failure and a right bundle bran...
متن کاملThe Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads
BACKGROUND Phrenic nerve stimulation (PNS) and high pacing thresholds (HPT) hinder biventricular stimulation in patients (pts) undergoing cardiac resynchronization therapy (CRT). A new quadripolar left ventricular (LV) lead (Quartet 1458Q, St. Jude Medical) with increased number of pacing configuration, might overcome this problem. METHODS All consecutive pts in whom a standard bipolar lead i...
متن کامل