Chronic Defibrillation Thresholds in Nonthoracotomy Implantable Defibrillator

نویسندگان

  • Ferdinand J. Venditti
  • David T. Martin
چکیده

Background To establish the chronic stability of defibrillation thresholds (DFTs) in a transvenous cardioverter/defibrillator (TCD) system, we studied 37 consecutive patients with TCD systems implanted for >6 months. Methods and Results DFT was measured by a step-down method at implant and 2 and 6 months later. The mean ejection fraction was 34.5±14.3%. Coronary artery disease with previous myocardial infarction was present in 31 patients. The mean DFT rose from 13.3 ±4.3 J at implant to 16.5 ±4.7 J at 2 months (P<.001) and 17.6±5.4 J at 6 months (P<.0001). ANOVA revealed a statistically significant rise in DFT over time (P<.0005). At 2 months, 25 patients had a rise in DFT, and 14 had a rise 25 J. The observed rise at 2 months persisted

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Electrode system influence on biphasic waveform defibrillation efficacy in humans.

BACKGROUND Several clinical studies have demonstrated a general superiority of biphasic waveform defibrillation compared with monophasic waveform defibrillation using epicardial lead systems. To test the breadth of utility of biphasic waveforms in humans, a prospective, randomized evaluation of defibrillation efficacy of monophasic and single capacitor biphasic waveform pulses was performed for...

متن کامل

Electrode System Influence on Biphasic

Background. Several clinical studies have demonstrated a general superiority of biphasic waveform defibrillation compared with monophasic waveform defibrillation using epicardial lead systems. To test the breadth of utility of biphasic waveforms in humans, a prospective, randomized evaluation of defibrillation efficacy of monophasic and single capacitor biphasic waveform pulses was performed fo...

متن کامل

Rise in chronic defibrillation energy requirements necessitating implantable defibrillator lead system revision.

The chronic defibrillation energy requirement (DER) is believed to remain clinically stable in patients with defibrillators. Six patients (two with an epicardial and four with a nonthoracotomy system) were identified with a rise in their chronic DER, which eliminated a 10-J safety margin, thus necessitating a defibrillator lead system revision. The mean increase in DER was 14.7 +/- 4 J and was ...

متن کامل

A permanent transvenous lead system for an implantable pacemaker cardioverter-defibrillator. Nonthoracotomy approach to implantation.

A transvenous lead system for implantable defibrillators would obviate a surgical thoracotomy and reduce the morbidity and mortality associated with implantation. We evaluated the clinical performance of a new nonthoracotomy lead system that included a defibrillation lead in the coronary sinus. At the time of defibrillator implantation, transvenous defibrillation leads were inserted percutaneou...

متن کامل

Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement

Evaluation of defibrillation threshold is a standard of care during implantation of implantable cardioverter defibrillator. High defibrillation thresholds are often encountered and pose a challenge to electrophysiologists to improve the defibrillation threshold. We describe a case series where defibrillation thresholds were improved after implanting a defibrillation lead in the azygos vein.