Comparison of an automated rapid plasma reagin (RPR) test with the conventional RPR card test in syphilis testing
نویسندگان
چکیده
OBJECTIVE We compared the automated non-treponemal reagin (rapid plasma reagin (RPR)) test with the conventional RPR card test for usefulness in clinical applications. SETTING A comparative study of laboratory methods using clinical specimens in a single institute. PARTICIPANTS A total of 112 serum samples including 59 Treponema pallidum particle agglutination (TPPA)-positive and 53 TPPA-negative specimens were evaluated. OUTCOME MEASURES HiSens Auto RPR LTIA (HBI, Anyang, Korea) was compared with Macro-Vue RPR Card Tests (Becton Dickinson BD Microbiology Systems, Sparks, Maryland, USA). Treponemal-specific tests were performed by Serodia TPPA assay (Fujirebio, Tokyo, Japan). The percentage agreement, κ value and overall sensitivity and specificity of the two RPR tests were compared. Seroconversion rates after treatment were also compared for each RPR test. RESULTS The percentage agreement between the two RPR tests was 78.6% (κ 0.565; 95% CI 0.422 to 0.709). Sensitivity and specificity of the automated RPR test relative to the TPPA test was 52.5% (95% CI 39.1% to 65.7%) and 94.3% (95% CI 84.3% to 98.8%), respectively, while the same values for the conventional RPR card test were 86.4% (95% CI 75% to 93.9%) and 94.3% (95% CI 84.3% to 98.8%), respectively. The conventional RPR card test showed overall higher positivity than the automated RPR test, whereas the automated RPR test showed higher seroconversion (43.5%, 10/23) than the conventional RPR card test (4.3%, 1/23) in treated patients. CONCLUSIONS The automated RPR test showed overall lower sensitivity than the conventional RPR test based on the treponemal test, but higher seroconversion after treatment. The automated RPR test could be used to monitor treatment response, especially in the reverse screening algorithm in syphilis testing.
منابع مشابه
Comparison of a new rapid plasma reagin card test with the standard rapid plasma reagin 18-mm circle card test and the venereal disease research laboratory slide test for serodiagnosis of syphilis.
The rapid plasma reagin (RPR) card test manufactured by Beckman Instruments, Inc., was compared, qualitatively and quantitatively, with the Venereal Disease Research Laboratory (VDRL) slide test and the standard RPR 18-mm circle card tests for the serodiagnosis of syphilis. Sera from 638 individuals were used in this study. Two pilot lots and two production lots of antigen were submitted by Bec...
متن کاملPractical Application of Quantitative HiSens Auto Rapid Plasma Reagin Latex Turbidimetric Immunoagglutination for Diagnosing Syphilis; Comparison Analysis between Rapid Plasma Reagin Latex Turbidimetric Immunoagglutination Test and Rapid Plasma Reagin Card Test
Background : The purpose of this study was to evaluate validity of quantitative RPR LTIA, HiSens Auto RPR LTIA (HBi Corp., South Korea) and to decide an adequate cutoff value for syphilis screening. Materials and Methods : A total of 549 serums or plasma specimens from patients were tested with RPR LTIA and RPR card tests. Degree of agreement between the two methods was analyzed, and validity o...
متن کاملEvaluation of a new rapid plasma reagin card test as a screening test for syphilis.
This study evaluates the American Dade (Biokit Laboratories) rapid plasma reagin (Dade RPR) card test, currently used in Spain for the diagnosis of syphilis, which has been recently released to the U.S. market. Used as a basis for comparison with the Dade card test were the 18-mm standard rapid plasma reagin (standard RPR) card test and the Venereal Disease Research Laboratory (VDRL) slide test...
متن کاملPerformance of the rapid plasma reagin and the rapid syphilis screening tests in the diagnosis of syphilis in field conditions in rural Africa.
OBJECTIVES To assess the rapid plasma reagin (RPR) test performance in the field and to evaluate a new rapid syphilis test (RST) as a primary screen for syphilis. METHODS 1325 women of reproductive age from rural communities in the Gambia were tested for syphilis seropositivity using a RPR 18 mm circle card and a RST strip. Within 1 week a repeat RPR and a TPHA test were carried out using sta...
متن کاملEDTA-treated plasma in the rapid plasma reagin card test and the toluidine red unheated serum test for serodiagnosis of syphilis.
Both the rapid plasma reagin (RPR) 18-mm circle card test and the toluidine red unheated serum test (TRUST) were designed to be more rapid than the Venereal Disease Research Laboratory slide test for the serodiagnosis of syphilis. The RPR card test, although originally designed as the teardrop card test for use with plasma, has received recognition from the Centers for Disease Control as a stan...
متن کامل