Investigation of the radiation dose from cone‐beam CT for image‐guided radiotherapy: A comparison of methodologies
نویسندگان
چکیده
Four methodologies were evaluated for quantifying kilovoltage cone-beam computed tomography (CBCT) dose: the Cone-Beam Dose Index (CBDI), IAEA Report 5 recommended methodology (IAEA), the AAPM Task Group 111 methodology (TG111), and the current dose metric; the Computed Tomography Dose Index (CTDI) on two commercial Varian cone-beam CT imaging systems; the Clinac iX On-Board Imager (OBI); and the TrueBeam X-ray Imaging system (XI). The TG111 methodology measured the highest overall dose (21.199 ± 0.035 mGy OBI and 22.420 ± 0.002 XI for pelvis imaging) due to the full scatter of the TG111 phantom and was within 5% of CTDI measurements taken using a full scatter TG111 phantom and 30-cm film strips. CBDI measured the second highest overall dose, within 10% of the TG111, with IAEA measuring the third highest dose. For head CBCT protocols, CBDI measured the highest dose, followed by IAEA. The CTDI method measured lowest across all scan modes highlighting its limitations for CBCT dosimetry. The XI imaging system delivered lower doses for head and thorax scan modes and similar doses to the OBI system for pelvis scan modes due to additional beam hardening filtration in the XI system. The TG111 method measured the highest dose in the center of a CBCT scan during image guidance procedures; however, CBDI provided a good approximation to TG111 with existing CTDI equipment and may be more applicable clinically.
منابع مشابه
The Impact of Residual Geometric Inaccuracies on Normal Organ Doses in Image Guided-Radiation Therapy of Prostate Cancer Using On-Board Kilovoltage Cone-Beam Computed Tomography
Introduction: The aim of this retrospective study was to evaluate the variations in delivered dose to the bladder, rectum, and femoral heads of prostate cancer patients during a course of treatment by image-guided radiation therapy (IGRT). Materials and Methods: Overall, 15 patients with prostate cancer were selected and. Each week, for each patient five consecutive cone beam computed tomograph...
متن کاملMethods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space
Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Materials and Methods: Compared to the traditional methods...
متن کاملComparative Study between Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) for Radiation Treatment Verifications
Introduction: Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) are the preferred tools of Image Guided Radiotherapy (IGRT) and Dose Guided Radiotherapy (DGRT) which have been used for Radiotherapy treatment verifications. As a result, the number of publications dealing with these two tools for radiation treatment verification has increased consi...
متن کاملUsability assessment of cone beam computed tomography with a full-fan mode bowtie filter compared to that with a half-fan mode bowtie filter
Background: In intensity modulated radiation therapy, cone beam computed tomography (CT) has been used to evaluate patients prior to treatment. This study conducted a comparative evaluation of the image reconstruction ability of the clinically used half-fan bowtie filter and the full-fan bowtie filter. Materals and Methods: A CT simulation marker was inserted inside a human phantom, and the pel...
متن کاملStandard edge detection algorithms versus conventional auto-contouring used for a three-dimensional rigid CT-CT matching
Background: To reduce uncertainties of patient positioning, the Computerized Tomography (CT) images acquired at the treatment planning time can be compared with those images obtained during radiation dose delivery. This can be followed during dose delivery procedure as Image Guided radiotherapy (IGRT) to verify the prescribed radiation dose delivery to the target as well as to monitor ...
متن کامل