Scaling Current and Energy with Body Weight: Requirements for the Transthoracic Ventricular Defibrillation of Calves
نویسنده
چکیده
To test the hypothesis that the effectiveness of a shock in achieving ventricular defibrillation is relatively independent of body weight if electrode diameter is proportional to the one-third power and current is proportional to the two-thirds power of weight, we studied defibrillation rates in 10 calves as they increased in weight. At 50 kg, each calf was subjected to 20 fibrillation-defibrillation episodes using 10.3-cm diameter electrodes and 32-amp, 4-msec rectangular pulses for defibrillation. Two days after the original study, each calf underwent 20 additional episodes involving 44-amp pulses. With the specified scaling of electrode diameter and pulse amplitude, the two studies were repeated at weight intervals of 25 kg as the animals grew. Six calves survived. In the study that started with 32-amp pulses, first-shock success values of 28%, 49%, 66%, 51% and 23% were found in the six surviving calves at 50, 75, 100, 125 and 150 kg, respectively. The corresponding values were 93%, 96%, 93%, 94% and 91% in the study that started with 44-amp pulses. While the results of the 32amp study fail to support our initial hypothesis, those obtained in the 44-amp current study appear compatible with the hypothesis.
منابع مشابه
Scaling current and energy with body weight: requirements for the transthoracic ventricular defibrillation of calves as they grow from 50 to 150 kg.
To test the hypothesis that the effectiveness of a shock in achieving ventricular defibrillation is relatively independent of body weight if electrode diameter is proportional to the one-third power and current is proportional to the two-thirds power of weight, we studied defibrillation rates in 10 calves as they increased weight. At 50 kg, each calf was subjected to 20 fibrillation-defibrillat...
متن کاملHypothesis for low-energy transthoracic defibrillation.
timely and informative. The Belfast group has a prodigious experience in out-of-hospital resuscitation and their prospectively collected data have been invaluable in determining energy requirements for defibrillation. The evidence indicating the need for defibrillators storing more than 400 Ws is, at present, inadequate and unconvincing (Lown et al., 1978). In addition to the views expressed by...
متن کاملAtrioventricular Nodal Reentrant Paroxysmal Tachycardia
amount of myocardium. Am J Cardiol 36: 37, 1975 5. Ewy GA: Defibrillation paddle electrodes. Proceedings of the Cardiac Defibrillation Conference, Purdue University, West Lafayette, Indiana, 1975, p 39 6. Ewy GA: Effectiveness of direct current defibrillation. Role of paddle electrode size. I1. Am Heart J 93: 674, 1977 7. Rush S, Lepeschkin E, Gregaritsch A: Current distribution from defibrilla...
متن کاملBRIEF REVIEW Determinants of successful defibrillation
Ventricular defibrillation by electric counter shock is the definitive treatment for cardiac arrest due to ventricular fibrillation (VF). Ventricular defibrillation is the fundamental cornerstone to a successful outcome. As most deaths from VF occur out of hospital, often in patients without pre-existent recognised cardiac disease, the major public health issue remains the widespread applicatio...
متن کاملElectrical dose for ventricular defibrillation of large and small animals using precordial electrodes.
Electrical ventricular defibrillation of heavy subjects (over 100 kg body weight) is uncommon for the human or any animal species. This paper reports trans-chest ventricular defibrillation of subjects ranging in weight from 2.3 to 340 kg using conventional defibrillation current (heavily damped sine wave) of 0.3-30 ms duration. It was found that a body weight-to-electrical-shock strength relati...
متن کامل