A novel method for vaginal cylinder treatment planning: a seamless transition to 3D brachytherapy

نویسندگان

  • Harish K Malhotra
  • Vincent Wu
  • Zhou Wang
  • Sachin Patil
چکیده

PURPOSE Standard treatment plan libraries are often used to ensure a quick turn-around time for vaginal cylinder treatments. Recently there is increasing interest in transitioning from conventional 2D radiograph based brachytherapy to 3D image based brachytherapy, which has resulted in a substantial increase in treatment planning time and decrease in patient through-put. We describe a novel technique that significantly reduces the treatment planning time for CT-based vaginal cylinder brachytherapy. MATERIAL AND METHODS Oncentra MasterPlan TPS allows multiple sets of data points to be classified as applicator points which has been harnessed in this method. The method relies on two hard anchor points: the first dwell position in a catheter and an applicator configuration specific dwell position as the plan origin and a soft anchor point beyond the last active dwell position to define the axis of the catheter. The spatial location of various data points on the applicator's surface and at 5 mm depth are stored in an Excel file that can easily be transferred into a patient CT data set using window operations and then used for treatment planning. The remainder of the treatment planning process remains unaffected. RESULTS The treatment plans generated on the Oncentra MasterPlan TPS using this novel method yielded results comparable to those generated on the Plato TPS using a standard treatment plan library in terms of treatment times, dwell weights and dwell times for a given optimization method and normalization points. Less than 2% difference was noticed between the treatment times generated between both systems. Using the above method, the entire planning process, including CT importing, catheter reconstruction, multiple data point definition, optimization and dose prescription, can be completed in ~5-10 minutes. CONCLUSION The proposed method allows a smooth and efficient transition to 3D CT based vaginal cylinder brachytherapy planning.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Caluculation of 3D Saptial Dose Distribution Around a Shielded Vaginal Cylinder with Iridium-192 Source Calculated by using Monte Carlo Code EGS4

According to our previous study on the dose distribution around a shielded vaginal cylinder with a high-intensity Ir source by using EGS4 code, the more detailed 3D-simulation model was established in this work. The results obtained by the 3D-simulation model were compared with those measured by ion chamber in previous work and evaluated by the existing treatment planning system. These results ...

متن کامل

Developing a Verification and Training Phantom for Gynecological Brachytherapy System

Introduction Dosimetric accuracy is a major issue in the quality assurance (QA) program for treatment planning systems (TPS). An important contribution to this process has been a proper dosimetry method to guarantee the accuracy of delivered dose to the tumor. In brachytherapy (BT) of gynecological (Gyn) cancer it is usual to insert a combination of tandem and ovoid applicators with a complicat...

متن کامل

Evaluation of dose distribution and dose gradient in brachytherapy cylindrical applicators using a dedicated Phantom for Iridium-192 and Cobalt-60 HDR sources

Introduction: A study was performed to evaluate radiation dose distribution and dose gradient around cylindrical applicators for high-dose-rate (HDR) brachytherapy systems with 192Ir, 60Co brachytherapy source applied for rectal and vaginal cancers treatments.   Materials and Methods: Two additional computed tomography (CT) based plans were generated using a ...

متن کامل

Modified Fletcher's 3-channel brachytherapy system with vaginal line source loading versus uterine tandem and vaginal cylinder system in Stage IIIA cervical cancer

PURPOSE The uterine tandem with open-ended vaginal cylinder is the most commonly used brachytherapy system for Federation Internationale de Gynecologie et d'Obstetrique (FIGO) Stage IIIA cervix cancer at the National Cancer Centre, Singapore. Without the 3-channel ovoid system, the dose to the parametrium is often compromised. In this study, a vaginal cylinder that could potentially be incorpor...

متن کامل

Effects of bladder distension on dose distribution of vaginal vault brachytherapy in patients with endometrial cancer

PURPOSE To investigate dosimetric effects of bladder distention on organs at risk (OARs) during treatment of endometrial cancer using 3D image-based planning of postoperative vaginal vault brachytherapy (BRT). MATERIAL AND METHODS Fifteen patients with early-stage endometrial cancer were studied, each undergoing adjuvant BRT of vaginal vault via 3.5 cm diameter cylinder. As treatment, 25 Gy i...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2012