Theoretical and experimental study on temperature elevation behind ribs caused by weakly focused ultrasound.

نویسندگان

  • Xiaozhou Liu
  • Chang Yin
  • Xiufen Gong
  • Wenwu Cao
چکیده

Temperature distribution in tissues behind ribs produced by weakly focused ultrasound had been calculated using Pennes bioheat equation and the validity of the theoretical model was experimentally confirmed in vitro using porcine live. We found that the position of the maximum temperature in tissue is strongly influenced by the distance between the transducer and ribs, while the gap between ribs is the determining factor for the maximum achievable temperature. Within the focal length, when the distance between the transducer and ribs increases, the maximum temperature increases and its position shifts away from the transducer. The rib width has little effect on the position of the maximum temperature but affects the achievable peak temperature. Our results provide useful information for treating liver cancers using ultrasound induced hyperthermia.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Temperature changes caused by the difference in the distance between the ultrasound transducer and bone during 1 mhz and 3 mhz continuous ultrasound: a phantom study

[Purpose] This study aimed to use a thermograph to observe temperature changes caused by different distances between an ultrasound transducer and bone during 1 MHz and 3 MHz continuous ultrasound emission on a phantom. [Materials and Methods] We observed the distribution of temperature elevations on a phantom consisting of pig ribs and tissue-mimicking material. One megahertz and 3 MHz ultrasou...

متن کامل

Quantitative assessment of acoustic intensity in the focused ultrasound field using hydrophone and infrared imaging.

With the popularity of ultrasound therapy in clinics, characterization of the acoustic field is important not only to the tolerability and efficiency of ablation, but also for treatment planning. A quantitative method was introduced to assess the intensity distribution of a focused ultrasound beam using a hydrophone and an infrared camera with no prior knowledge of the acoustic and thermal para...

متن کامل

Focus Splitting Associated with Propagation of Focused Ultrasound through the Rib Cage.

The effect of focus splitting after propagation of focused ultrasound through a rib cage is investigated theoretically. It is shown that the mechanism of this effect is caused by the interference of waves from two or more spatially separated sources, such as intercostal spaces. Analytical estimates of the parameters of splitting are obtained, i.e., the number of foci, their amplitudes, diameter...

متن کامل

Experimental Temperature Monitoring and Coagulation Detection using Ultrasound-Stimulated Acoustic Emission

Abstract – Due to the high variability of tissue effects during sonication, temperature monitoring is one of the most crucial components for accurate thermal treatment of tissues with Focused Ultrasound (FUS) and other thermotherapy devices. Since the technique of ultrasoundstimulated acoustic emission (USAE) depends on the acoustic and mechanical properties, which both vary with temperature, i...

متن کامل

اندازه‌گیری ضخامت چربی زیرپوستی در تصاویر مادون قرمز در لیپوساکشن: گزارش کوتاه

Background: Currently, there are many techniques to measure subcutaneous body fat but these methods have many limitations. In this study, we tried to provide a clustering algorithm to measure the thickness of subcutaneous fat in thermography images. Methods: For the detection of subcutaneous adipose tissue in the midline area (from pubis to the xiphoid process), imaging takes place in the righ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Ultrasound in medicine & biology

دوره 36 10  شماره 

صفحات  -

تاریخ انتشار 2010