Reliability of the EchoMRI-Infant System for Water and Fat Measurements in Newborns
نویسندگان
چکیده
OBJECTIVE The precision and accuracy of a quantitative magnetic resonance (EchoMRI Infants) system in newborns were determined. METHODS Canola oil and drinking water phantoms (increments of 10 g to 1.9 kg) were scanned four times. Instrument reproducibility was assessed from three scans (within 10 minutes) in 42 healthy term newborns (12-70 hours post birth). Instrument precision was determined from the coefficient of variation (CV) of repeated scans for total water, lean mass, and fat measures for newborns and the mean difference between weight and measurement for phantoms. In newborns, the system accuracy for total body water (TBW) was tested against deuterium dilution (D2 O). RESULTS In phantoms, the repeatability and accuracy of fat and water measurements increased as the weight of oil and water increased. TBW was overestimated in amounts >200 g. In newborns weighing 3.14 kg, fat, lean mass, and TBW were 0.52 kg (16.48%), 2.28 kg, and 2.40 kg, respectively. EchoMRI's reproducibility (CV) was 3.27%, 1.83%, and 1.34% for total body fat, lean mass, and TBW, respectively. EchoMRI-TBW values did not differ from D2 O; mean difference, -1.95 ± 6.76%, P = 0.387; mean bias (limits of agreement), 0.046 kg (-0.30 to 0.39 kg). CONCLUSIONS The EchoMRI Infants system's precision and accuracy for total body fat and lean mass are better than established techniques and equivalent to D2 O for TBW in phantoms and newborns.
منابع مشابه
Calibration and validation of EchoMRI™ whole body composition analysis based on chemical analysis of piglets, in comparison with the same for DXA.
A study was conducted to appraise a new EchoMRI™ device for body composition analysis (BCA) of infants and to compare it with dual energy X-ray absorptiometry (DXA), using chemical analysis as a reference method.The calibration part of the study included cross-validation comparisons between EchoMRI™ measurements of awake, anesthetized and dead piglets of the calibration set. It also included co...
متن کاملEvaluation of a quantitative magnetic resonance imaging system for whole body composition analysis in rodents.
We evaluated the EchoMRI-900 combination rat and mouse quantitative magnetic resonance (QMR) body composition method in comparison to traditional whole-body chemical carcass composition analysis (CCA) for measurements of fat and fat-free mass in rodents. Live and postmortem (PM) QMR fat and lean mass measurements were obtained for lean, obese and outbred strains of rats and mice, and compared w...
متن کاملThe use of bioelectrical impedance analysis in newborns. The need for standardization.
Rapid growth occurs during fetal and neonatal development. The duration and quality of intrauterine development affect neonate body composition. Weight is routinely used to assess and monitor the nutritional status of newborns. Any weight loss or gain in hospitalized newborn infants may reflect shifts in water balance rather than changes in body mass, and have strong implications for investigat...
متن کاملBrown Adipose Tissue Quantification in Human Neonates Using Water-Fat Separated MRI
There is a major resurgence of interest in brown adipose tissue (BAT) biology, particularly regarding its determinants and consequences in newborns and infants. Reliable methods for non-invasive BAT measurement in human infants have yet to be demonstrated. The current study first validates methods for quantitative BAT imaging of rodents post mortem followed by BAT excision and re-imaging of exc...
متن کاملAssessment of Body Composition Using Bioelectrical Impedance Analysis in Preterm Neonates Receiving Intensive Care.
AIM Evaluation of bioelectrical impedance values and body composition during early infancy in groups of preterm newborns and full-term newborns. MATERIAL AND METHODS A total of 38 newborns was enrolled in the study: 26 very low birth weight preterm newborns with the mean birth weight of 1236 g (SD: 161) as the study group and 12 term newborns with birth weight of 2500-4000 g as the control gr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 25 شماره
صفحات -
تاریخ انتشار 2017