Innovations in intra-aortic balloon pump management: computer modem technology.

نویسنده

  • J M Ingram
چکیده

The computer modem technology being developed by Datascope Corporation for the purpose of off-site intra-aortic balloon pump (IABP) monitoring was clinically evaluated in 30 patients requiring IABP support. The Datascope System 95 model IABP with built-in modem was used on all patients. Remote communications via a personal computer with modem were conducted under both routine and emergency settings. The Datascope PC IABP software was evaluated for its user compatibility, efficiency, diagnostic capability and overall usefulness as a clinical tool. During the eight month evaluation period, 87 remote communications were conducted for both routine and emergency IABP evaluation checks. Adjustments were recommended on 22 occasions relevant to balloon timing, trigger mode selection and augmentation volume settings. Eight communications were initiated in emergency settings due to a variety of patient conditions. Emergency intervention was successful in diagnosing and resolving critical situations including atrial arrhythmias, pacemaker timing, low cardiac output syndrome, loss of trigger source, catheter malpositioning and poor augmentation. The diagnostic capabilities and efficient means of data collection by the computer software provide the clinician with a valuable tool for routine IABP clinical monitoring, as well as emergency problem resolution.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Evaluation of Survival in Patients Who Need Intra Aortic Balloon Pump (IABP) after Cardiac Surgery

Introduction: Insome cases, assist devices are required to ensure an adequate cardiac output after cardiac surgery. Intra-aortic balloon pump (IABP) is the most commonly used cardiac assist device, which provides first-line support for patients with heart failure. The aim of this study was to determine factors affecting the mortality rate of patients receiving IABP. Materials & Methods:  In thi...

متن کامل

Biventricular pacing in conjunction with epicardial atrial pacing adversely affects timing of intra-aortic balloon pump.

A case of accidental triggering of an intra-aortic balloon pump during systole is presented. The patient had a cardiac resynchronization therapy device in place preoperatively for heart failure. A temporary epicardial atrial pacing wire was used during separation from cardiopulmonary bypass for rate control. An intra-aortic balloon pump was necessary for separation from bypass. Although the pac...

متن کامل

Management of Intra-Aortic Balloon Pump Rupture and Entrapment.

While the intra-aortic balloon pump (IABP) itself has many useful and life-saving indications, it has a high complication rate, up to 30% in some studies. This report discusses the surgical removal and subsequent vascular repair in a patient with a ruptured and subsequently entrapped IABP and reviews the literature on the presentation and management of this rare but major complication.

متن کامل

Intra-aortic balloon pump in patients with cardiogenic shock complicating myocardial infarction: a systematic review and meta-analysis of randomized trials (protocol)

BACKGROUND Cardiogenic shock is the leading cause of death in patients with acute myocardial infarction. Despite significant advancements in health technology and research, hospital mortality approaches 50%. The intra-aortic balloon pump is a mechanical hemodynamic assist device that has been used for over 40 years in the management of patients with cardiogenic shock. A recent randomized trial ...

متن کامل

A novel technique for intra-aortic balloon positioning in the intensive care unit.

Proper positioning of the intra-aortic balloon catheter is very important as an improperly placed balloon can cause problems. A balloon placed too high can block off the arch vessels, whereas one placed too low can block the renal and splanchnic vessels. We propose a simple and reliable technique to properly position the intra-aortic balloon pump (IABP) catheter without the use of any sophistic...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 26 3  شماره 

صفحات  -

تاریخ انتشار 1994