Rapid microarray detection of DNA and proteins in microliter volumes with surface plasmon resonance imaging measurements.
نویسندگان
چکیده
A four-chamber microfluidic biochip is fabricated for the rapid detection of multiple proteins and nucleic acids from microliter volume samples with the technique of surface plasmon resonance imaging (SPRI). The 18 mm × 18 mm biochip consists of four 3 μL microfluidic chambers attached to an SF10 glass substrate, each of which contains three individually addressable SPRI gold thin film microarray elements. The 12-element (4 × 3) SPRI microarray consists of gold thin film spots (1 mm(2) area; 45 nm thickness), each in individually addressable 0.5 μL volume microchannels. Microarrays of single-stranded DNA and RNA (ssDNA and ssRNA, respectively) are fabricated by either chemical and/or enzymatic attachment reactions in these microchannels; the SPRI microarrays are then used to detect femtomole amounts (nanomolar concentrations) of DNA and proteins (ssDNA binding protein and thrombin via aptamer-protein bioaffinity interactions). Microarrays of ssRNA microarray elements are also used for the ultrasensitive detection of zeptomole amounts (femtomolar concentrations) of DNA via the technique of RNase H-amplified SPRI. Enzymatic removal of ssRNA from the surface due to the hybridization adsorption of target ssDNA is detected as a reflectivity decrease in the SPR imaging measurements. The observed reflectivity loss is proportional to the log of the target ssDNA concentration with a detection limit of 10 fM or 30 zeptomoles (18 000 molecules). This enzymatic amplified ssDNA detection method is not limited by diffusion of ssDNA to the interface, and thus is extremely fast, requiring only 200 s in the microliter volume format.
منابع مشابه
On-chip synthesis of protein microarrays from DNA microarrays via coupled in vitro transcription and translation for surface plasmon resonance imaging biosensor applications.
Protein microarrays are fabricated from double-stranded DNA (dsDNA) microarrays by a one-step, multiplexed enzymatic synthesis in an on-chip microfluidic format and then employed for antibody biosensing measurements with surface plasmon resonance imaging (SPRI). A microarray of dsDNA elements (denoted as generator elements) that encode either a His-tagged green fluorescent protein (GFP) or a Hi...
متن کاملSurface Enzyme Chemistries for Ultrasensitive Microarray Biosensing with SPR Imaging
The sensitivity and selectivity of surface plasmon resonance imaging (SPRI) biosensing with nucleic acid microarrays can be greatly enhanced by exploiting various nucleic acid ligases, nucleases, and polymerases that manipulate the surface-bound DNA and RNA. We describe here various examples from each of these different classes of surface enzyme chemistries that have been incorporated into nove...
متن کاملBiofunctionalization of Gold Nanorods: A Comparative Study on Conjugation Methods for Fabrication of Nanobiosensors
Gold Nanorods have promised variety of applications in biomedicine and biosensing. As a fruitful candidate for early detection and imaging, these plasmonic nanoparticles have been utilized for diagnostic applications of interest. However, prior to design and fabricate SPR-based nanobiosensors, the type and nature of conjugation with biomolecules would be of utmost importance. Herein, four strat...
متن کاملUltrasensitive microarray detection of short RNA sequences with enzymatically modified nanoparticles and surface plasmon resonance imaging measurements.
A novel multiplexed method for short RNA detection that employs an enzymatic capture reaction onto DNA-modified silica nanoparticles (SiNPs) followed by nanoparticle-enhanced surface plasmon resonance imaging (SPRI) is demonstrated. SiNPs functionalized with 5'-phosphorylated single stranded DNA (ssDNA) are used with T4 RNA ligase to capture various short 20-24 base single-stranded RNA (ssRNA) ...
متن کاملCombined Lab-on-a-chip and Microarray Approach for Biomolecular Interaction Sensing Using Surface Plasmon Resonance Imaging
Surface plasmon resonance imaging (SPR) is a well-established label-free detection technique for real-time biomolecular interaction measurements. An integrated LOC sensing system with fluidic control for sample movement to specific locations on microarray surface in combination with SPR imaging is demonstrated by the measurements of human IgG and anti-IgG interactions from 24 patterned regions.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Langmuir : the ACS journal of surfaces and colloids
دوره 27 10 شماره
صفحات -