Technical Note: Intercomparison of formaldehyde measurements at the atmosphere simulation chamber SAPHIR
نویسندگان
چکیده
The atmosphere simulation chamber SAPHIR at the Research Centre Jülich was used to test the suitability of state-of-the-art analytical instruments for the measurement of gas-phase formaldehyde (HCHO) in air. Five analyzers based on four different sensing principles were deployed: a differential optical absorption spectrometer (DOAS), cartridges for 2,4-dinitrophenylhydrazine (DNPH) derivatization followed by off-line high pressure liquid chromatography (HPLC) analysis, two different types of commercially available wet chemical sensors based on Hantzsch fluorimetry, and a proton-transfer-reaction mass spectrometer (PTRMS). A new optimized mode of operation was used for the PTR-MS instrument which significantly enhanced its performance for online HCHO detection at low absolute humidities. The instruments were challenged with typical ambient levels of HCHO ranging from zero to several ppb. Synthetic air of high purity and particulate-filtered ambient air were used as sample matrices in the atmosphere simulation chamber onto which HCHO was spiked under varying levels of humidity and ozone. Measurements were compared to mixing ratios calculated from the chamber volume and the known amount of HCHO injected into the chamber; measurements were also compared between the different instruments. The formal and blind intercomparison exercise was conducted Correspondence to: T. Brauers ([email protected]) under the control of an independent referee. A number of analytical problems associated with the experimental set-up and with individual instruments were identified, the overall agreement between the methods was fair.
منابع مشابه
Intercomparison of Hantzsch and fiber-laser-induced-fluorescence formaldehyde measurements
Two gas-phase formaldehyde (HCHO) measurement techniques, a modified commercial wet-chemical instrument based on Hantzsch fluorimetry and a custombuilt instrument based on fiber laser-induced fluorescence (FILIF), were deployed at the atmospheric simulation chamber SAPHIR (Simulation of Atmospheric PHotochemistry In a large Reaction Chamber) to compare the instruments’ performances under a rang...
متن کاملIsotope effect in the formation of H2 from H2CO studied at the atmospheric simulation chamber SAPHIR
Formaldehyde of known, near-natural isotopic composition was photolyzed in the SAPHIR atmosphere simulation chamber under ambient conditions. The isotopic composition of the product H2 was used to determine the isotope effects in formaldehyde photolysis. The experiments are sensitive to the molecular photolysis channel, and the radical channel has only an indirect effect and cannot be effective...
متن کاملSAPHIR actinometry
Actinometric measurements of NO2 photolysis frequencies in the atmosphere simulation chamber SAPHIR B. Bohn, F. Rohrer, T. Brauers, and A. Wahner Institut für Chemie und Dynamik der Geosphäre II: Troposphäre, Forschungszentrum Jülich, 52425 Jülich, Germany Received: 25 October 2004 – Accepted: 2 December 2004 – Published: 8 December 2004 Correspondence to: B. Bohn ([email protected]) © 2004 ...
متن کاملModel-aided radiometric determination of photolysis frequencies in a sunlit atmosphere simulation chamber
In this work diurnal and seasonal variations of mean photolysis frequencies for the atmosphere simulation chamber SAPHIR at Forschungszentrum Jülich are calculated. SAPHIR has a complex construction with UV permeable teflon walls allowing natural sunlight to enter the reactor volume. The calculations are based on external measurements of solar spectral actinic flux and a model considering the t...
متن کاملActinometric measurements of NO2 photolysis frequencies in the atmosphere simulation chamber SAPHIR
The simulation chamber SAPHIR at Forschungszentrum Jülich has UV permeable teflon walls facilitating atmospheric photochemistry studies under the influence of natural sunlight. Because the internal radiation field is strongly affected by construction elements, we use external, radiometric measurements of spectral actinic flux and a model to calculate mean photolysis frequencies for the chamber ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -