Accuracy and Longevity of an Implantable Continuous Glucose Sensor in the PRECISE Study: A 180-Day, Prospective, Multicenter, Pivotal Trial.

نویسندگان

  • Jort Kropff
  • Pratik Choudhary
  • Sankalpa Neupane
  • Katharine Barnard
  • Steve C Bain
  • Christoph Kapitza
  • Thomas Forst
  • Manuela Link
  • Andrew Dehennis
  • J Hans DeVries
چکیده

OBJECTIVE It is known that continuous glucose monitoring (CGM) systems can lower mean glucose compared with episodic self-monitoring of blood glucose. Implantable CGM systems may provide additional benefits. RESEARCH DESIGN AND METHODS We studied the Eversense (Senseonics Inc.) implantable CGM sensor in 71 participants aged 18 years and older with type 1 and type 2 diabetes in a 180-day multinational, multicenter pivotal trial. Participants used the CGM system at home and in the clinic. CGM accuracy was assessed during eight in-clinic visits with the mean absolute relative difference (MARD) for venous reference glucose values >4.2 mmol/L as the primary end point. Secondary end points included Clarke Error Grid Analysis and alarm performance. The primary safety outcome was device-related serious adverse events. This trial is registered with ClinicalTrials.gov, number NCT02154126. RESULTS The MARD value against reference glucose values >4.2 mmol/L was 11.1% (95% CI 10.5, 11.7). Clarke Error Grid Analysis showed 99.2% of samples in the clinically acceptable error zones A and B. Eighty-one percent of hypoglycemic events were detected by the CGM system within 30 min. No device-related serious adverse events occurred during the study. CONCLUSIONS Our results indicate the safety and accuracy of this new type of implantable CGM system and support it as an alternative for transcutaneous CGM.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Prospective Multicenter Evaluation of the Accuracy of a Novel Implanted Continuous Glucose Sensor: PRECISE II

BACKGROUND Persistent use of real-time continuous glucose monitoring (CGM) improves diabetes control in individuals with type 1 diabetes (T1D) and type 2 diabetes (T2D). METHODS PRECISE II was a nonrandomized, blinded, prospective, single-arm, multicenter study that evaluated the accuracy and safety of the implantable Eversense CGM system among adult participants with T1D and T2D (NCT02647905...

متن کامل

Analysis: Continuous glucose monitoring during intensive insulin therapy.

Results of the Normoglycemia in Intensive Care Evaluation and Survival Using Glucose Algorithm Regulation (NICE-SUGAR) trial, intensive insulin therapy (IIT), and use of a continuous glucose sensor in intensive care units (ICU) were analyzed. The NICE-SUGAR trial was unable to determine if optimal intensive insulin therapy decreases mortality. Continuous glucose monitoring (CGM) technology has ...

متن کامل

Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

OBJECTIVE The purpose of this study was to compare the accuracy of measurements of glucose in interstitial fluid made with the FreeStyle Navigator Continuous Glucose Monitoring System with Yellow Springs Instrument laboratory reference measurements of venous blood glucose. RESEARCH DESIGN AND METHODS Fifty-eight subjects with type 1 diabetes, aged 18-64 years, were enrolled in a multicenter, ...

متن کامل

Accuracy of the SEVEN continuous glucose monitoring system: comparison with frequently sampled venous glucose measurements.

BACKGROUND The purpose of this study was to compare the accuracy of measurements obtained from the DexCom SEVEN system with Yellow Springs Instrument (YSI) laboratory measurements of venous blood glucose. METHODS Seventy-two subjects with insulin-requiring diabetes, aged 18-71, were enrolled in a multicenter, prospective single-arm study. All participants wore the SEVEN continuous glucose mon...

متن کامل

Continuous Glucose Sensor Survival and Accuracy Over 14 Consecutive Days

Continuous glucose monitoring (CGM) technology provides real-time glucose data and the ability to achieve improved glycemic control; however, widespread adoption has been limited (1). Among the challenges to CGMuptake are cost and the inconvenience of changing the sensor. In this prospective, multicenter study, theDexcomG4PLATINUMsystemwas analyzed over 14 days of continuous sensor wear. Each s...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Diabetes care

دوره 40 1  شماره 

صفحات  -

تاریخ انتشار 2017