Evaluation of the Bilichek transcutaneous bilirubinometer in the Chinese new-borns
نویسندگان
چکیده
Objective: To evaluate the relation between transcutaneous bilirubin (TcB) by BiliChek and total serum bilirubin (TSB) in the Chinese neonates compared with another transcutaneous Bilirubinometer (VanHou). Methods: TcB measurements were performed on the forehead by Bilichek and VanHou. Total serum bilirubin (TSB) was determined within 5 minutes after TcB measurement. The correlation and agreement between the two devices and TSB were assessed. Results: A total of 172 babies were recruited. The mean TSB (± SD) was 11.1 ± 4.6 mg/dl, and the mean TcB (± SD) was respectively 13.4 ± 3.5 mg/dl (VanHou), 12.3 ± 4.5 mg/dl (Bilichek). When the TSB was lower than 9 mg/dl, the difference of VanHou TcB and TSB was much higher than that of Bilichek (3.51 vs. 1.28 mg/dl, p<0.01). When the TSB was 9 ~ 12 mg/dl, the bias of TcB and TSB was 2.31 mg/dl (VanHou) vs. 1.39 mg/dl (Bilichek) (p<0.01). The correlation between the TcB (Bilichek) and TSB was a little higher than the VanHou (0.94 vs. 0.92). Bilichek had a greater area under the ROC curve than VanHou when a threshold value of TSB was chosen at 9, 12 or 15 mg/dl. The Bland-Altman analysis demonstrated that TcB tended to overestimate the TSB, and compared with VanHou, Bilichek had a better agreement with the TSB. Conclusions: Bilichek has a good correlation with TSB. Furthermore, when TSB was lower than 12 mg/dl, Bilichek is superior to the VanHou obviously.
منابع مشابه
Evaluation of a point-of-care transcutaneous bilirubinometer in Chinese neonates at an accident and emergency department.
OBJECTIVE To evaluate the use of a point-of-care transcutaneous bilirubinometer, JM-103 Minolta, for estimation of the serum bilirubin level in the management of neonatal jaundice in term or near-term Chinese neonates. DESIGN Prospective correlation study. SETTING Accident and Emergency Department of a regional hospital in Hong Kong. PATIENTS All term or near-term Chinese neonates aged 3 ...
متن کاملمقایسه میانگین بیلی روبین اندازه گیری شده از طریق بیلی روبینومتر با میانگین بیلی روبین سرم در نوزادان زرد
Introduction: Hyperbilirubinemia is a common benign problem in neonates. Nonetheless, if untreated, severe hyperbilirubinemia is potentially neurotoxic. The standard method for diagnosis of hyperbilirubinemia is determination of serum bilirubin levels in the laboratory that requires a blood sample which is both painful and stressful for neonates and entails cost and time. The objective of thi...
متن کاملTranscutaneous bilirubinometry: an evaluation.
The transcutaneous bilirubinometer was evaluated in 60 term and 10 preterm infants. A significant correlation was found between the transcutaneous index and the total serum bilirubin concentration for both term and preterm infants. The reliability of the transcutaneous bilirubinometer as a screening method was confirmed, and index criteria for serum bilirubin analysis have been suggested for te...
متن کاملA Comparison between Transcutaneous Bilirubin (TcB) and Total Serum Bilirubin (TSB) Measurements in Term Neonates
Background: Transcutaneous bilirubinometry (TCB) is a simple method for estimating bilirubin levels in neonates. This method is noninvasive, quick, and painless. We aimed to compare serum and cutaneous bilirubin measurements in term neonates.Method: In this descriptive cross-sectional study, 200 neonates with icter with birth weights of at least 2500 grams were studied. TCB was measured using a...
متن کاملTranscutaneous bilirubinometry in Malay, Chinese and Indian term neonates.
The transcutaneous (Tc) bilirubinometer was evaluated in 105 jaundiced neonates, comprising 38 Malays, 37 Chinese and 30 Indians, who had not been treated with phototherapy or exchange transfusion. Tc bilirubin index and serum bilirubin concentration correlated at statistically significant levels in all the three racial groups. Unlike the Chinese and Malay babies, the action levels, using Tc bi...