Tourniquet safety in lower leg applications.

نویسندگان

  • James A McEwen
  • Deborah L Kelly
  • Theda Jardanowski
  • Kevin Inkpen
چکیده

PURPOSE To reduce the chance of injury due to pneumatic tourniquet use, the minimum cuff pressure required to maintain a bloodless field should be used. The purpose of this study was to find out if Limb Occlusion Pressure (LOP--the cuff pressure required to occlude arterial flow) is lower with a wide contoured cuff than with a standard width cylindrical cuff at the calf, if cuff pressures based on measured LOP will be lower than the typical 250 mmHg used in lower leg cuffs, and if a new automatic LOP measurement method gives the same results as the standard Doppler stethoscope method. SAMPLE 16 adult volunteers were tested in a controlled laboratory setting, and 53 clinical cases were reviewed at two centers. DESIGN Repeated measures comparison of LOP on volunteers with the two different cuffs and measurement methods, and review of clinical cases. RESULTS LOP was lower with the wide cuff on all volunteers (mean reduction 20 mmHg, SD 8.6, range 5-35, p < 0.001). The average difference of 1.2 mmHg between Doppler and automatic LOP readings was not significant (p = 0.43). Based on the volunteer results, using LOP plus a safety margin of 40, 60, or 80 mmHg (for LOP < 130, 131-190, or 190+ respectively) with a standard width cylindrical cuff will lead to an average cuff pressure of 223 mmHg (range 170-299, SD 36), 11% lower than typical practice and up to 80 mmHg (32%) lower on some patients. Using a wide, contoured cuff should further reduce cuff pressures to an average of 195 mmHg (range 160-280, SD 33), 22% lower than typical practice and a reduction of up to 90 mmHg (36%). At two clinics, the wide cuff maintained a bloodless field in 48 out of 53 cases (91%) when used at 200 mmHg. CONCLUSIONS Using a wide, contoured cuff at the calf should reduce required cuff pressures compared to a standard cuff. Setting cuff pressure based on LOP should further reduce cuff pressures for most patients compared to typically used pressures. With continued development, the new automatic method may become a viable alternative to the Doppler method and may make LOP measurement more practical in the clinical setting.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The use of lower tourniquet inflation pressures in extremity surgery facilitated by curved and wide tourniquets and an integrated cuff inflation system.

Use of the lowest possible cuff inflation pressure should minimize the pathogenic effects of compression beneath the pneumatic tourniquet. Curved tourniquets (designed to fit conically shaped limbs) and wider tourniquets were associated with significantly lower arterial occlusion pressure (AOP) than standard, straight tourniquets on the arms and legs of 26 normal volunteers. These tourniquets w...

متن کامل

The Effect of the Silicone Ring Tourniquet and Standard Pneumatic Tourniquet on the Motor Nerve Conduction, Pain and Grip Strength in Healthy Volunteers

  Background: The pneumatic tourniquet (PT) is routinely used in upper and lower limb operations by most orthopaedic surgeons. The silicone ring tourniquet (SRT) was introduced in clinical practice the last decade. Clinical as well comparative studies in volunteers concerning its safety and efficacy have been published. The aim of this study was to investigate the postoperative effect of the si...

متن کامل

An evaluation of two tourniquet systems for the control of prehospital lower limb hemorrhage.

BACKGROUND Hemorrhage remains the main cause of preventable death on the modern battlefield. As Improvised Explosive Devices in Afghanistan become increasingly powerful, more proximal limb injuries occur. Significant concerns now exist about the ability of the windlass tourniquet to control distal hemorrhage after mid-thigh application. To evaluate the efficacy of the Combat Application Tourniq...

متن کامل

Tourniquet cuff pressure and nerve injury in knee arthroplasty in a bloodless field

BACKGROUND AND PURPOSE Tourniquet-related nerve injuries remain a concern in orthopedic surgery. The cuff pressures used today are generally lower, and therefore a decreasing incidence of peripheral nerve injuries might also be expected. However, there have been few neurophysiological studies describing the outcome after bloodless field surgery. We describe the results of neurophysiological exa...

متن کامل

Metabolic effects of tourniquet ischaemia studied by nuclear magnetic resonance spectroscopy.

A model of tourniquet ischaemia was developed in the hind limb of the rat, and the metabolic changes that occurred in the calf muscles were monitored by the non-invasive technique of phosphorus-31 nuclear magnetic resonance spectroscopy. During ischaemia the intramyocellular pH became acidic as the level of phosphocreatine declined and that of inorganic phosphate rose. Phosphocreatine was no lo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Orthopedic nursing

دوره 21 5  شماره 

صفحات  -

تاریخ انتشار 2002