Objective assessment, repeatability, and agreement of shoulder ROM with a 3D gyroscope
نویسندگان
چکیده
BACKGROUND Assessment of shoulder mobility is essential for diagnosis and clinical follow-up of shoulder diseases. Only a few highly sophisticated instruments for objective measurements of shoulder mobility are available. The recently introduced DynaPort MiniMod TriGyro ShoulderTest-System (DP) was validated earlier in laboratory trials. We aimed to assess the precision (repeatability) and agreement of this instrument in human subjects, as compared to the conventional goniometer. METHODS The DP is a small, light-weight, three-dimensional gyroscope that can be fixed on the distal upper arm, recording shoulder abduction, flexion, and rotation. Twenty-one subjects (42 shoulders) were included for analysis. Two subsequent assessments of the same subject with a 30-minute delay in testing of each shoulder were performed with the DP in two directions (flexion and abduction), and simultaneously correlated with the measurements of a conventional goniometer. All assessments were performed by one observer. Repeatability for each method was determined and compared as the statistical variance between two repeated measurements. Agreement was illustrated by Bland-Altman-Plots with 95% limits of agreement. Statistical analysis was performed with a linear mixed regression model. Variance for repeated measurements by the same method was also estimated and compared with the likelihood-ratio test. RESULTS Evaluation of abduction showed significantly better repeatability for the DP compared to the conventional goniometer (error variance: DP = 0.89, goniometer = 8.58, p = 0.025). No significant differences were found for flexion (DP = 1.52, goniometer = 5.94, p = 0.09). Agreement assessment was performed for flexion for mean differences of 0.27° with 95% limit of agreement ranging from -7.97° to 8.51°. For abduction, the mean differences were 1.19° with a 95% limit of agreement ranging from -9.07° to 11.46°. CONCLUSION In summary, DP demonstrated a high precision even higher than the conventional goniometer. Agreement between both methods is acceptable, with possible deviations of up to greater than 10°. Therefore, static measurements with DP are more precise than conventional goniometer measurements. These results are promising for routine clinical use of the DP.
منابع مشابه
تاثیر شش هفته تمرینات مقاومتی با کش بر میزان درد مفصل و دامنه حرکتی مردان ورزشکار مبتلا به سندروم گیرافتادگی شانه
Background: Among all shoulder injuries, shoulder impingement syndrome (SIS) is more common and is the most common cause for pain and stiffness in the shoulder area. The study purpose was to investigate the effect of six weeks resistance exercises with elastic-band on joint pain and range of motion (ROM) in athlete men with shoulder impingement syndrome. Methods: This was a semi experimental re...
متن کاملAmbulatory assessment of shoulder abduction strength curve using a single wearable inertial sensor.
The aim of the present article was to assess the reliability of strength curves as determined from tridimensional linear accelerations and angular velocities measured by a single inertial measurement unit (IMU) fixed on the upper arm during a shoulder abduction movement performed holding a 1 kg dumbbell in the hand. Within-subject repeatability of the task was assessed on 45 subjects performing...
متن کاملEffectiveness of Action Potential Simulation and Transcutaneous Electrical Nerve Stimulation on Pain and Function of Patients With Chronic Mechanical Shoulder Impairment
Purpose: Rehabilitation and physiotherapy have a major role in reducing the pain and increasing the functional ability in patients with shoulder joint pain. This study aimed to evaluate the effect of Action Potential Simulation (APS) and Transcutaneous Electrical Nerve Stimulation (TENS) in patients with Chronic Mechanical Shoulder Pain (CMSP). Methods: In this quasi-experimental stu...
متن کاملStrength and Range of Motion of Internal and External Rotator Muscles in Volleyball Players With and Without Uneven Shoulders
Objective: Performing repetitive overhead movements puts stress on the shoulder joint, leading to uneven shoulders and pain in the shoulder joint. The purpose of this study was to evaluate and compare the strength and Range of Motion (ROM) of internal and external rotator muscles in volleyball players with and without uneven shoulders. Methods: This is a descriptive and causal-comparative stud...
متن کاملCase study: a novel biomechanical approach for evaluating extended body armor systems.
To combat the devastating effects of improvised explosive devices (IEDs), body armor that provides extended coverage has been developed. However, this extended coverage increases the armor's weight and may restrict movement. Throughout this case study, a novel technique to assess several armor systems was investigated. Four soldiers performed shoulder and trunk movements while wearing each of t...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 14 شماره
صفحات -
تاریخ انتشار 2013