Transthoracic resistance in human defibrillation. Influence of body weight, chest size, serial shocks, paddle size and paddle contact pressure.

نویسندگان

  • R E Kerber
  • J Grayzel
  • R Hoyt
  • M Marcus
  • J Kennedy
چکیده

Successful defibrillation depends on delivery of adequate electrical current to the heart; one of the major determinants of current flow is transthoracic resistance (TTR). To study the factors influencing TTR, we prospectively collected data from 44 patients undergoing emergency defibrillation. Shocks of 94-450 J delivered energy were administered from specially calibrated Datascope defibrillators that displayed peak current flow, thereby permitting determination of TTR. Shocks were applied from standard (8.5-cm diameter) or large (13 cm) paddles placed anteriorly and laterally. First-shock TTR ranged from 15-143 omega. There was a weak correlation between TTR and body weight (r = 0.45, p less than 0.05) and a stronger correlation between TTR and chest width (r = 0.80, p less than 0.01). Twenty-three patients who were defibrillated using standard 8.5-cm paddles had a mean TTR of 67 +/- 36 omega (+/- SD), whereas 21 patients who received shocks using paddle pairs with at least one large (13 cm) paddle had a 21% lower TTR of 53 +/- 24 omega (p = 0.05, unpaired t test). Ten patients received first and second shocks at the same energy level; TTR declined only 8%, from 52 +/- 19 to 48 +/- 16 omega (p less than 0.01, paired t test). In closed chest dogs, shocks were administered using a spring apparatus that regulated paddle contact pressure against the thorax. Firmer contact pressure caused TTR to decrease 25%, from 48 +/- 22 to 36 +/- 17 omega (p less than 0.01, paired t test). Thus, human TTR varies widely and is related most closely to chest size. TTR declines only slightly with a second shock at the same energy level. More substantial reductions in TTR and declines only slightly with a second shock at the same energy level. More substantial reductions in TTR and increases in current flow can be achieved by using large paddles and applying firm paddle contact pressure.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Atrioventricular Nodal Reentrant Paroxysmal Tachycardia

amount of myocardium. Am J Cardiol 36: 37, 1975 5. Ewy GA: Defibrillation paddle electrodes. Proceedings of the Cardiac Defibrillation Conference, Purdue University, West Lafayette, Indiana, 1975, p 39 6. Ewy GA: Effectiveness of direct current defibrillation. Role of paddle electrode size. I1. Am Heart J 93: 674, 1977 7. Rush S, Lepeschkin E, Gregaritsch A: Current distribution from defibrilla...

متن کامل

Determinants of Intracardiac Current in Defibrillation

Defibrillation success depends upon the amount and distribution of intracardiac current (IC), because only a portion of total current flow between the paddles actually traverses the heart. We evaluated two determinants of IC: paddle size and shock energy. Studies were performed in 11 anesthetized dogs using a damped-sine wave form defibrillator. Intramyocardial electrode pairs were implanted su...

متن کامل

Transthoracic defibrillation: effect of sternotomy on chest impedance.

The purpose of this study was to determine the effect of sternotomy on transthoracic impedance, a major determinant of current flow and defibrillation success. Transthoracic impedance was determined by using a validated test-pulse technique that does not require actual shocks. Seventeen patients undergoing median sternotomy were studied prospectively. Transthoracic impedance was determined befo...

متن کامل

System for open-chest, multidirectional electrical defibrillation

Introduction: Cardiomyocytes are more sensitive to stimulatory electrical fields when the latter are applied longitudinally to the cell major axis. In the whole heart, cells have different spatial orientations, which may limit the effectiveness of conventional electrical defibrillation (i.e., shock delivery in a single direction). This article describes the constructive aspects of a portable sy...

متن کامل

Cardiac Pacemaker Battery Discharge After External Electrical Cardioversion for Broad QRS Complex Tachycardia

External electrical cardioversion or defibrillation may be necessary in patients with implanted cardiac pacemaker (PM) or implantable cardioverter defibrillator (ICD). Sudden discharge of high electrical energy employed in direct current (DC) transthoracic countershock may damage the PM/ICD system resulting in a series of possible device malfunctions. For this reason, when defibrillation or car...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 63 3  شماره 

صفحات  -

تاریخ انتشار 1981