Standardization of high-sensitivity immunoassays for measurement of C-reactive protein; II: Two approaches for assessing commutability of a reference material.
نویسندگان
چکیده
BACKGROUND We evaluated the commutability of a proposed reference material (PRM), with a formulation based on dilution of Certified Reference Material 470 (CRM470), for 24 high-sensitivity C-reactive protein (hsCRP) methods. We also investigated whether calibration by use of PRM was effective in harmonizing results. METHODS A set of 40 native clinical samples was measured along with PRM and 3 dilutions of PRM. We used weighted least-squares polynomial regression (WLS/PR) to perform comparisons between all method combinations and to calculate normalized residuals for the PRM. The PRM was considered noncommutable if any of the normalized residuals for a method pair was >2. Correspondence analysis (CA) was used to explore the multidimensional relationships between methods and samples to evaluate if the PRM had properties similar to native clinical samples. Clinical sample results from the methods for which PRM was commutable were recalibrated based on the PRM results, and ANOVA was used to estimate the CVs before and after recalibration. RESULTS After omitting data for 9 methods because of poor precision or procedural flaws, we used data from the 15 remaining methods to evaluate commutability. Using both WLS/PR and CA we found that PRM was noncommutable with 1 method. We found modest improvement in total and among-method CVs when PRM was used to harmonize the results from the 14 methods for which it was commutable. CONCLUSIONS A PRM with a formulation based on dilution of CRM470 was commutable with native clinical samples for 14 of 15 hsCRP methods that had acceptable precision. For those methods the use of PRM may contribute to improved harmonization of results for native clinical samples.
منابع مشابه
Standardization of immunoassays for measurement of high-sensitivity C-reactive protein. Phase I: evaluation of secondary reference materials.
BACKGROUND Inflammation contributes to the development and progression of atherosclerosis, and C-reactive protein (CRP) can be used as a marker to assess risk for cardiovascular diseases. As variability among existing high-sensitivity CRP (hsCRP) assays can lead to misclassification of patients and hamper implementation of population-based medical decision points, standardization of hsCRP assay...
متن کاملWhy commutability matters.
A fundamental goal of laboratory medicine is that results for patients’ samples will be comparable independent of the medical laboratory that produced the results. Routine measurement procedures of acceptable analytical specificity that have calibration traceable to the same higherorder reference material or reference measurement procedure should produce numerical values for clinical samples th...
متن کاملCounterpoint: Food and Drug Administration guidance for C-reactive protein assays: matching claims with performance data.
LM, et al. CDC/AHA workshop on markers of inflammation and cardiovascular disease: application to clinical and public health practice: report from the laboratory science discussion group. Circulation 2004;110:e545–9. 4. Ridker PM, Wilson PW, Grundy SM. Should C-reactive protein be added to metabolic syndrome and to assessment of global cardiovascular risk? Circulation 2004;109:2818–25. 5. US Fo...
متن کاملCommutability of the ERM-DA470k reference material for two assays measuring serum albumin using immunochemical principles.
The standardization of measurements is a high-priority in Laboratory Medicine; its purpose being to achieve closer comparability of results for the same analyte obtained using different commercial systems (1). The promotion of result traceability to available higher-order reference measurement procedures and reference materials is the recommended approach (2). A major prerequisite for guarantee...
متن کاملToward standardization of cardiac troponin I measurements part II: assessing commutability of candidate reference materials and harmonization of cardiac troponin I assays.
BACKGROUND Cardiac tropoin I (cTnI) measurements show an approximately 20- to 40-fold difference between assays, and better standardization and harmonization are needed. Toward this goal, the AACC cTnI Standardization Committee collaborated with the National Institute of Standards and Technology (NIST) in an earlier study to select 2 candidate reference materials (cRMs). METHODS Two troponin ...