In vitro assessment of Function Graded (FG) artificial Hip joint stem in terms of bone/cement stresses: 3D Finite Element (FE) study

نویسندگان

  • Fawzi F Al-Jassir
  • H Fouad
  • Othaman Y Alothman
چکیده

BACKGROUND Stress shielding in the cemented hip prosthesis occurs due to the mismatching in the mechanical properties of metallic stem and bone. This mismatching in properties is considered as one of the main reasons for implant loosening. Therefore, a new stem material in orthopedic surgery is still required. In the present study, 3D finite element modeling is used for evaluating the artificial hip joint stem that is made of Function Graded (FG) material in terms of joint stress distributions and stem length. METHOD 3D finite element models of different stems made of two types of FG materials and traditional stems made of Cobalt Chromium alloy (CoCrMo) and Titanium alloy (Ti) were developed using the ANSYS Code. The effects on the total artificial hip joint stresses (Shear stress and Von Mises stresses at bone cement, Von Mises stresses at bone and stem) due to using the proposed FG materials stems were investigated. The effects on the total artificial hip joint system stresses due to using different stem lengths were investigated. RESULTS Using FG stem (with low stiffness at stem distal end and high stiffness at its proximal end) resulted in a significant reduction in shear stress at the bone cement/stem interface. Also, the Von Mises stresses at the bone cement and stem decrease significantly when using FG material instead of CoCrMo and Ti alloy. The stresses' distribution along the bone cement length when using FG material was found to be more uniform along the whole bone cement compared with other stem materials. These more uniform stresses will help in the reduction of the artificial hip joint loosening rate and improve its short and long term performance. CONCLUSION FE results showed that using FG stem increases the resultant stresses at the femur bone (reduces stress shielding) compared to metallic stem. The results showed that the stem length has significant effects on the resultant shear and Von Mises stresses at bone, stem and bone cement for all types of stem materials.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of Various Functionally Graded Femoral Prostheses by Finite Element Analysis

This study is focused on finite element analysis of a model comprising femur into which a femoral component of a total hip replacement was implanted. The considered prosthesis is fabricated from a functionally graded material (FGM) comprising a layer of a titanium alloy bonded to a layer of hydroxyapatite. The elastic modulus of the FGM was adjusted in the radial, longitudinal, and longitudinal...

متن کامل

Finite Element Analysis of Different Hip Implant Designs along with Femur under Static Loading Conditions

Background: The hip joint is the largest joint after the knee, which gives stability to the whole human structure. The hip joint consists of a femoral head which articulates with the acetabulum. Due to age and wear between the joints, these joints need to be replaced with implants which can function just as a natural joint. Since the early 19th century, the hip joint arthroplasty has evolved, a...

متن کامل

Effects of prosthesis surface roughness on the failure process of cemented hip implants after stem-cement debonding.

Retrieval studies suggest that the loosening process of the cemented femoral components of total hip arthroplasties is initiated by failure of the bond between the prosthesis and the cement mantle. Finite element (FE) analyses have demonstrated that stem-cement debonding has stress-producing effects in the cement mantle. High interface friction, which corresponds to a degree of surface roughnes...

متن کامل

Buckling Behaviors of Symmetric and Antisymmetric Functionally Graded Beams

The present study investigates buckling characteristics of both nonlinear symmetric power and sigmoid functionally graded (FG) beams. The volume fractions of metal and ceramic are assumed to be distributed through a beam thickness by the sigmoid-law distribution (S-FGM), and the symmetric power function (SP-FGM). These functions have smooth variation of properties across the boundary rather tha...

متن کامل

Viscoelastic analysis of stress distribution in balanced and unbalanced adhesively bonded single-lap joints with functionally graded adherends under the Reddy model

In this study, shear and peel stress distributions in the viscoelastic adhesive layer of a single-lap joint (SLJ) with functionally graded (FG) adherends are investigated. The study focuses on the effect of different adherend profiles and material composition on the time-dependent stress concentration/distribution in balanced and unbalanced SLJs. For this purpose, the Reddy model is applied to ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2013